background image

53

────────────────────────────────────────────────────

7.4 Commands Specific to the 3157

────────────────────────────────────────────────────

:SYSTem:OPTion:CCHange

Sets the current changeability in the TEST state.

:SYSTem:OPTion:CCHange?

Queries the current changeability in the TEST state.

 Syntax 

:SYSTem:OPTion:CCHange

_<data>

 <data> 

Numerical data in NR1 format between 0 and 1

 Function 

Sets the current changeability in the TEST state.
0: Not changeable
1: Changeable
The numerical value can be in NRf format, but any digits after the decimal
point will be rounded.

 

Example

 

Transmission

 

:SYSTem:OPTion:CCHange 1

The current changeability in the TEST state is set to "Changeable."

 Error 

If <data> is other than numerical value described above, an execution error
occurs.
The execution of this command in a state other than the following states
causes an execution error:

READY state

Optional function setting screen

 Syntax 

:SYSTem:OPTion:CCHange?

 Function 

Returns the setting for the current changeability in the TEST state as a
numerical value in NR1 format.

 <data> 

Numerical data in NR1 format between 0 and 1

 

Example

 

Response

 

If headers are on 

:SYSTEM:OPTION:CCHANGE 1

If headers are off 

1

 Error 

If the response message is longer than 300 bytes, a query error is generated.

Содержание 9593-02

Страница 1: ...9593 02 Instruction Manual RS 232C INTERFACE EN Oct 2018 Revised edition 4 9593A985 04 18 10H For 3157 AC GROUNDING HiTESTER...

Страница 2: ......

Страница 3: ...Connection to the Printer 13 4 1 9442 PRINTER 13 4 2 Setting the 9442 PRINTER Communication Conditions 14 4 3 Connecting the Printer 15 Chapter 5 Connection to the 3155 17 Chapter 6 Operation 19 6 1 S...

Страница 4: ...Explanations 32 7 3 Particular Commands 33 7 4 Commands Specific to the 3157 35 7 5 Response Format for Queries as Numerical Value 67 7 6 Initialization Items 68 Chapter 8 Sample Programs 69 8 1 Samp...

Страница 5: ...you for purchasing this HIOKI 9593 02 RS 232C INTERFACE To get the maximum performance from the unit please read this manual first and keep this at hand This Instruction Manual provides information a...

Страница 6: ...ii Introduction...

Страница 7: ...nal Appearance and Accessories When the unit is delivered check and make sure that it has not been damaged in transit If the unit is damaged or fails to operate according to the specifications contact...

Страница 8: ...ng the unit preferably use the original packing If this is not available use the following procedure 1 Wrap the unit in plastic sheeting 2 After wrapping cushioning material around the unit pack it in...

Страница 9: ...o the new ones 2 Always be sure to secure the RS 232C cable to the 9593 02 unit by tightening up the fixing screws provided 3 Program messages sent just after the power has been turned on are executed...

Страница 10: ...opening of the removed unit CAUTION When inserting in the interface hold the metal plate Directly touching the board may cause static electricity and lead to damage of the instrument Using the wrist s...

Страница 11: ...pter 2 Overview 2 1 Introduction to the RS 232C Interface By connecting the 9593 02 RS 232C INTERFACE to the 3157 AC GROUNDING HiTESTER it is possible to control all the functions of the main unit exc...

Страница 12: ...ich the momentary OUT settings are disabled even though the momentary OUT settings have been set 2 The beeper sound can be turned on and off 3 The unit can be reset 4 Test results can be printed on th...

Страница 13: ...RS 232C interface connector pin assignments D subminiature 25 pin female NOTE Signal assignments and explanation Connector Dsub pin number Circuit Description RS 232C CCITT 1 AA FG 101 Protective Grou...

Страница 14: ...D subminiature 25pin male 9593 02 BA TxD CF DCD CA RTS CB CTS 2 3 4 5 6 7 8 8 2 3 1 4 5 7 BB RxD AB GND CD DTR BB RxD CA RTS AB GND CF DCD BA TxD CC DSR CB CTS D subminiature 9pin female PC AT 2 20 6...

Страница 15: ...s set to Space to indicate to the controller that the empty buffer capacity is low Processing of data in the buffer continues and when the receiving buffer is less than 25 full CA RTS is set to Mark t...

Страница 16: ...10 2 3 Specifications...

Страница 17: ...mote state releasing keys Press the 0ADJ key while holding down the SHIFT key to release the remote state by the RS 232C and resume the local state However this key operation is disabled if the RS 232...

Страница 18: ...nication condition setting switches These are used to set the communication conditions of the 3157 unit on the RS 232C bus For how to set these switches refer to Section 5 1 Setting the RS 232C Commun...

Страница 19: ...he test values can be printed out The 9442 is shipped with the function settings for use with the HIOKI 3166 CLAMP ON POWER HiTESTER Before using always change the settings of the DIP switches For det...

Страница 20: ...Baud rate See the table below Baud rate bps Switch No 19200 5 6 7 8 OFF ON ON OFF 4 2 Setting the 9442 PRINTER Communication Conditions The 9442 is shipped with the function settings for use with the...

Страница 21: ...unit with an RS 232C interface removed If you should wish to use the unit after removing an interface fit a blank panel over the opening of the removed unit 4 3 Connecting the Printer To connect the 9...

Страница 22: ...number of test data Not printed for OFF HIOKI 3157 GROUNDING HiTESTER Model name This is printed if the test data count function is not used or if the number of test data is 1 DATA Number of test data...

Страница 23: ...ables testing when connected with 3155 3155 sends command to 3157 to start testing and receives test results when the 3157 testing is complete The test results can be saved and printed together with t...

Страница 24: ...tion cable as specified below Settings 1 Leave power OFF for both 3155 and 3157 while connecting each RS 232C connector with the RS 232C cable 2 Turn the power ON for both 3155 and 3157 3 Set up 3157...

Страница 25: ...ication condition setting switches on the RS 232C panel to set the communication conditions On dispatch from the factory this address is initially set to 00000000 If you change the communication condi...

Страница 26: ...m messages are command messages or query messages Command messages are orders for controls of the 3157 such as for making unit settings or for reset or the like Example TIMER_ data Command message whi...

Страница 27: ...n this manual the short form is written in upper case letters and then this is continued in lower case letters so as to constitute the long form Either of these forms will be accepted during operation...

Страница 28: ...Examples CONFigure CURRent MEASure RESistance etc Particular header A header beggining with an asterisk to indicate that it is a particular command Examples RST etc 2 Response message Headers in resp...

Страница 29: ...ats identified as NR1 NR2 and NR3 and each of these can appear as either a signed number or an unsigned number Unsigned numbers are taken as positive Further if the accuracy of a numerical value excee...

Страница 30: ...en it is desired to set out several messages on a single line Example UNIT_OHM UPPER_ON CONFIGURE RUPPER_0 100 When messages are combined in this way if a syntax error occurs all subsequent messages u...

Страница 31: ...he directory structure of UNIX or MSDOS and until it is cleared the analysis of following commands is performed by deeming them to be preceded by the current path which has been curtailed in the inter...

Страница 32: ...on again The 3157 has an output queue of 300 bytes capacity If the response messages overflow this limit of 300 bytes a query error is generated and the output queue is cleared The 3157 has an input...

Страница 33: ...is outside the set range The designated data value is not acceptable Bit 3 DDE Device dependent error When a command cannot be executed due to some cause other than a command error a query error or an...

Страница 34: ...s register 0 ESR0 bit assignments Bit 7 Unused Bit 6 Unused Bit 5 Unused Bit 4 Unused Bit 3 EOM Test completed Bit 2 LFAIL Below lower limit of comparator Bit 1 UFAIL Above upper limit of comparator B...

Страница 35: ...mmand Explanation Ref page CLS Clears event register 31 ESR Queries standard event status register SESR 31 IDN Queries device ID 31 RST Device initialization 32 TST Queries the result of the self test...

Страница 36: ...maximum test value voltage 40 CONFigure VUPPer Queries the maximum test value voltage 40 ESR0 Queries event status register 0 40 HEADer Enables and disables headers for the response messages 41 HEADer...

Страница 37: ...Queries the output current frequency 55 SYSTem OPTion HOLD Sets the hold function 56 SYSTem OPTion HOLD Queries the hold function 56 SYSTem OPTion LOWer Sets the minimum test value 57 SYSTem OPTion L...

Страница 38: ...ommand Note Specifies precautions to be taken when using the command Example These are simple examples of the use of the command Note that all transmission messages are expressed in a short form Error...

Страница 39: ...er this command a command error occurs Syntax ESR Function Returns the contents of the standard event status register SESR as a numerical value in NR1 format between 0 and 255 and then clears standard...

Страница 40: ...yntax RST Function Resets the 3157 The items which are reset are listed below Error If the data parameters are set after this command a command error occurs Syntax TST Function Performs the self test...

Страница 41: ...zero adjustment function is turned on Error If data is other than character data and numerical value described above a command error occurs The execution of this command in a state other than the READ...

Страница 42: ...ue is set and set to OFF returns OFF When the endless timer function is set returns When the endless timer function is not set and the test time is set to OFF returns OFF Example Response If headers a...

Страница 43: ...CONFigure DATA_ data data Numerical data in NR1 format between 1 and 99 Function Sets the number of test data The numerical value can be in NRf format but any digits after the decimal point will be r...

Страница 44: ...resistance and when the minimum test value setting is not disabled and when ON is selected in the ON OFF setting for the minimum test value Example Transmission CONFigure RLOWer 0 000 The minimum tes...

Страница 45: ...e set to be expressed in OHM resistance and when ON is selected in the ON OFF setting for the maximum test value Example Transmission CONFigure RUPPer 0 100 The maximum test value resistance is set to...

Страница 46: ...sabled and when ON is selected in the ON OFF setting for the test time Example Transmission CONFigure TIMer 60 0 The test time is set to 60 0 s Error If data is other than numerical value described ab...

Страница 47: ...voltage and when the minimum test value setting is not disabled and when ON is selected in the ON OFF setting for the minimum test value Example Transmission CONFigure VLOWer 0 00 The minimum test va...

Страница 48: ...ing for the maximum test value Example Transmission CONFigure VUPPer 2 50 The maximum test value voltage is set to 2 50 V Error If data is other than numerical value described above an execution error...

Страница 49: ...ssages When powering on data is initially set to OFF Example Transmission HEADer ON Headers are prefixed to response messages Error If data is other than character data described above an execution er...

Страница 50: ...entry is possible or not depends on the state of the unit For more information see the instruction manual for the 3157 Example Transmission KEY 0 2 The right arrow key entry is set Error If data 1 an...

Страница 51: ...ent Function Returns the measured current value as a numerical value in NR2 format data Numerical data in NR2 format between 0 0 and 35 0 Note The measured current value is returned in the TEST state...

Страница 52: ...ge is longer than 300 bytes a query error is generated Syntax MEASure TIMer Function Returns the test time elapsed as a numerical value in NR2 format data character data Numerical data in NR2 format b...

Страница 53: ...red value and result as measured current value measured voltage value test time elapsed screening result in order If the unit of the maximum and minimum test values is set to OHM resistance the measur...

Страница 54: ...the measured voltage value is required query using the MEASure VOLTage command until the next test starts In the case of an overflow for the measured resistance O F is returned The test time elapsed i...

Страница 55: ...minimum test value is not set returns When the minimum test value is set and set to OFF returns OFF When the endless timer function is set returns When the endless timer function is not set and the t...

Страница 56: ...e optional function setting screen Note The STARt command works in a state in which the momentary OUT settings are disabled even though the momentary OUT settings have been set Example Transmission ST...

Страница 57: ...result returns to the READY state In the save screen returns to the READY state without saving In the load screen returns to the READY state without loading Example Transmission STOP Error If the data...

Страница 58: ...nded Example Transmission SYSTem OPTion BUZZer 3 The buzzer is set to ON at screening OFF at error Error If data is other than numerical value described above an execution error occurs The execution o...

Страница 59: ...SYSTem OPTion CCHange 1 The current changeability in the TEST state is set to Changeable Error If data is other than numerical value described above an execution error occurs The execution of this com...

Страница 60: ...the test Example Transmission SYSTem OPTion CDATa 10 The maximum number of test data is set to 10 Error If data is other than numerical value described above an execution error occurs If a smaller val...

Страница 61: ...test data is not reflected in the test Example Transmission SYSTem OPTion COUNt 1 The test data count function is set Error If data is other than numerical value described above an execution error occ...

Страница 62: ...is not reflected in the test Example Transmission SYSTem OPTion ENDLess 1 The endless timer function is set Error If data is other than numerical value described above an execution error occurs The e...

Страница 63: ...PTion FREQuency 1 The output current frequency is set to 60 Hz Error If data is other than numerical value described above an execution error occurs The execution of this command in a state other than...

Страница 64: ...n HOLD 1 The hold function is set to Held Error If data is other than numerical value described above an execution error occurs The execution of this command in a state other than the following states...

Страница 65: ...gh the momentary OUT settings have been set Example Transmission SYSTem OPTion LOWer 1 The minimum test value is set Error If data is other than numerical value described above an execution error occu...

Страница 66: ...Example Transmission SYSTem OPTion MOMentary 1 The momentary OUT function is set Error If data is other than numerical value described above an execution error occurs In the continuous test mode sett...

Страница 67: ...unded Example Transmission SYSTem OPTion PFHold 3 The PASS FAIL hold function is set to PASS held FAIL not held Error If data is other than numerical value described above an execution error occurs Th...

Страница 68: ...will be rounded Example Transmission SYSTem OPTion PRINter 0 The printer output function is set to Not used Error If data is other than numerical value described above an execution error occurs The ex...

Страница 69: ...OUT function is automatically set to Not set Example Transmission SYSTem OPTion TMODe 2 The test mode is set to Continuous test mode Error If data is other than numerical value described above an exec...

Страница 70: ...ON The test time is set to ON Error If data is other than character data and numerical value described above a command error occurs The execution of this command in a state other than the following st...

Страница 71: ...o VOLT voltage the maximum and minimum test values voltage settings are reflected in the test Example Transmission UNIT OHM The unit of the maximum and minimum test values is set to OHM Error If data...

Страница 72: ...a is other than character data and numerical value described above a command error occurs The execution of this command in a state other than the following states causes an execution error READY state...

Страница 73: ...Two or three digits in NR2 format 2 Maximum test value voltage minimum test value voltage and measured voltage value Three digits in NR2 format 3 Maximum test value resistance minimum test value resi...

Страница 74: ...vice specific functions ranges etc 1 Output queue Input buffer Event registers 2 Current path Headers on off Measurement resister 7 6 Initialization Items The following table shows which items are ini...

Страница 75: ...1 Sample program Flowchart 1 Basic settings and testing This program sets the 3157 to the following test values Output current value 25 0 A Maximum test value 0 100 Test time 60 0 s It carries out a s...

Страница 76: ...70 8 1 Sample program Flowchart...

Страница 77: ...t current value 25 0 A 10 0 A 25 0 A 15 0 A 10 0 A Maximum test value 0 100 1 00 V 0 100 1 50 V 0 100 Test time 60 0 s 10 0 s 5 0 s OFF 5 0 s 2 Saving the basic settings This program sets the 3157 to...

Страница 78: ...72 8 1 Sample program Flowchart Yes Y No N 3 Inquiry on and the loading of the contents of memory This program loads the contents of Setting memory in the 3157 after making this inquiry...

Страница 79: ...y processing Check the test Check the test finished Check the test Test results processing 4 Testing at five positions and outputting of a text data file After test settings five tests are made and th...

Страница 80: ...the result on the screen Program List 100 OPEN COM1 9600 N 8 1 FOR RANDOM AS 1 110 STATE CK1 120 PRINT 1 STAT 130 LINE INPUT 1 A 140 IF A READY THEN GOTO SETTING 150 PRINT 1 STOP 160 GOTO STATE CK1 1...

Страница 81: ...results 410 Close the RS 232C circuit file Sample display Press any key to start 25 0 0 020 60 0 PASS 2 Saving the basic settings This program sets the 3157 to the following test values and then save...

Страница 82: ...TE CK5 640 SETTING5 650 PRINT 1 CONF CURR 10 0 660 PRINT 1 UNIT OHM 670 PRINT 1 UPP ON 680 PRINT 1 CONF RUPP 0 100 690 PRINT 1 TIM ON 700 PRINT 1 CONF TIM 5 0 710 PRINT 1 MEM SAVE 5 720 CLOSE 730 END...

Страница 83: ...ILE CK 180 PRINT 1 MEM FILE 1 190 LINE INPUT 1 B 200 PRINT B 210 KEY IN 220 INPUT Load Y N C 230 IF C Y THEN GOTO FILE LOAD 240 IF C N THEN GOTO EXIT1 250 GOTO KEY IN 260 FILE LOAD 270 PRINT 1 MEM LOA...

Страница 84: ...TIM ON 280 PRINT 1 CONF TIM 5 0 290 TESTING 300 GOSUB KEY IN 310 PRINT 1 STAR 320 GOSUB STATE TEST 330 GOSUB STATE TESTEND 340 GOSUB RESULT 350 IF N 5 THEN GOTO EXIT1 360 GOTO TESTING 370 EXIT1 380 P...

Страница 85: ...m that the 3157 is in the TEST state 530 580 Check that a test has finished 590 680 Display the test results and outputs them to a file Sample display Drive A File XXX Press any key to start 1 25 1 0...

Страница 86: ...80 8 2 Sample programs...

Страница 87: ...a time and read the responses individually When you want to read them in all at once try doing so by putting them all on one line separated by the message separator character The response message to...

Страница 88: ...82...

Страница 89: ......

Страница 90: ......

Страница 91: ......

Страница 92: ......

Отзывы: