51
────────────────────────────────────────────────────
7.4 Commands Specific to the 3157
────────────────────────────────────────────────────
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
Unused Unused Unused Unused Unused Overrun
error
Framing
error
Parity
error
RS-232C communication condition errors register
:STOP
■
Performs forcible ending of a test and releases the hold state.
:SYSTem:ERRor?
■
Queries RS-232C communication condition errors.
Syntax
:STOP
Function
In the TEST state (test in progress), performs forcible ending of a test.
When retaining the test result, returns to the READY state.
In the save screen, returns to the READY state without saving.
In the load screen, returns to the READY state without loading.
Example
Transmission
:STOP
Error
If the data parameters are set after this command, a command error occurs.
Syntax
:ERRor?
Function
Returns the value of RS-232C communication condition errors as a numerical
value in NR1 format from 0 to 7, and then clears RS-232C communication
condition errors.
No header is prefixed to the response message.
<data>
Numerical data in NR1 format between 0 and 7
Example
Response
4
An overrun error has occurred.
Error
If the response message is longer than 300 bytes, a query error is generated.