Operation Manual                                                                                                                                                                                                                                         
 

BinTra

c Operator’s Manual Ver. 3.18 Rev. 1.01 

Part Number MAN-000004-2

 

Setup (Version 3.0 and higher) 

 
The Bin LEDs indicate configuration options as being enabled (solid on) or disabled (flashing).  

If not using any of the 

below options, ensure that all bins are showing as disabled (flashing) when 

 is shown on the display.

 

 
Bin A 

– Configures BinTrac Indicator as a Remote Display. 

 

A Remote Display is hardwired to a BinTrac Indicator for remotely viewing weight data.  

Bin B 

– Enable ASCII Serial Communications Command Set.  (See Below) 

BinTrac Indicator or BinTrac Remote Display will transmit weight data based on received commands.  Enable 
this feature when interfacing unit to a PC or serial type device. 

Bin C 

– Enable Weight Broadcast. 

ASCII weight data will be serially broadcast approximately once ever second. 

Bin D 

– Enable communications to peripheral devices. 

Must be enabled when BinTrac Indicator is connected to a BinTrac Remote Display and/or a HouseLink PW. 
 

1.  Press the BIN key to select the desired configuration option. 
2.  Use the UPPER 

 or LOWER 

 keys to enable or disable options. 

 

Segmented display: 

 

 

 

Serial Data Command Set 

 

Poll 
Command 

Description and  
Data Record Format (fixed length)

 

P   

Current weight of first enabled bin 
+/-wwwwwcrlf 

PP  

Current weight of all four bins 
+/-wwwww,+/-wwwww,+/-wwwww,+/-wwwwwcrlf 

SnnP  

Current weight of first enabled bin with matching BinTrac Station ID 
+/-wwwwwcrlf 
*Also works with F, FF and U, UU commands 

SnnPP crlf 

Current weight of all four bins with matching BinTrac Station ID 
+/-wwwww,+/-wwwww,+/-wwwww,+/-wwwwwcrlf 
*Also works with F, FF and U, UU commands 

Fcr 

Single Bin Fill: yy,mm,dd,hh,mm,+/-wwwwwcrlf 

Ucr 

Single Bin Usage: yy,mm,dd,hh,mm,+/-wwwwwcrlf 

FFcr 

Multiple Bin Fill: 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 

UUcr 

Multiple Bin Usage: 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 
yy,mm,dd,hh,mm,+/-wwwwwcrlf 

Note:  All commands must be terminated by CR.   Latency between characters within a multiple character        
command (including CR), cannot exceed 250msec. 

– bin weight 

 

 

 

yy,mm,dd,hh,mm 

– Year, Month, Day, Hour, Minutes 

– Station ID number 

 

 

CR 

– Carriage Return (Hx0D). 

LF 

– Line Feed (Hx0A). 

 

***The above Serial settings are for BinTrac Indicator software versions 3.08 and earlier. Please refer to the 
HouseLink HL-10S Serial Interface Manual for any BinTrac Indicator software versions 3.15 or higher.

 

Содержание BINTRaC BT200

Страница 1: ...em Operation Manual BT200 BT260 U S Patent No 7 980 129 Patent No 8 581 122 Patent No 8 853 566 U S Patent No 9 310 243 Patent No 9 651 413 Patent No 10 082 421 Canada Patent No 2 822 294 Korea Patent...

Страница 2: ...0 and higher 7 Serial Data Command Set 7 Setup Mode Settings 8 Internal Mode 11 Accessing the Internal mode 11 Navigating Internal mode 11 Selecting an Option in Internal mode 11 Internal 11 Totalizin...

Страница 3: ...s that could result from voids holes in the material Smart Summing Box One Smart Summing Box per bin communicates the current reading on the leg brackets to the BinTrac Indicator BinTrac Power Supply...

Страница 4: ...r indicators will illuminate showing their current status If a bin indicator is on steady it is selected as being monitored If the indicator is flashing that bin is not being monitored 2 Press the BIN...

Страница 5: ...alert level is set to one of the upper 8 indicators top half of the bin alert is configured as a high level alert The output relay will trigger when the indicated bin level is at or above this point 6...

Страница 6: ...elease the BIN key The display will then move to the bin selection as show below Segmented display Ver 3 17 below most BT200s Press and hold the BIN key until is displayed then release the BIN key The...

Страница 7: ...egmented display Serial Data Command Set Poll Command Description and Data Record Format fixed length P Current weight of first enabled bin wwwwwcrlf PP Current weight of all four bins wwwww wwwww www...

Страница 8: ...just this value appropriately IMPORTANT DO NOT change this setting from the default of 3 000 if you are using HerdStar load cells 1 Press the BIN key to select the desired bin 2 Use the UPPER key to i...

Страница 9: ...ts the weight of a full bin This is for calibration of the LED bar graph level The value dictates at what net weight the bar graph will display completely full all 16 LEDs lit 1 Press the BIN key to s...

Страница 10: ...Segmented display Software Version Version 3 0 and higher Displays the BinTrac programmed software version number This number may be required if technical help is needed 1 Press the BIN key to see the...

Страница 11: ...ultiple Summing Boxes V3 03 BIN D Change transmitted weight values from fixed 5 digits to fixed 6 digits Totalizing Multiple Bins For applications which require more than one Smart Summing Box on a bi...

Страница 12: ...ss the BIN key to display and allow editing of programmed value Select the desired bin 2 Use the UPPER key to increase the pulse weight by 1 3 Use the LOWER key to decrease the pulse weight by 1 Segme...

Страница 13: ...Indicator through a series of LED tests Segmented display Year Year is set at the time of production and should NOT be changed Segmented display Month Month is set at the time of production and shoul...

Страница 14: ...corded fill events 5 Press BIN to return to Auto Sequence mode or unit will start auto sequencing after 5 minutes of no activity no keystrokes Display 24 Hour Usage This procedure allows you to view t...

Страница 15: ...mming Boxes to ensure it has not been damaged 5 Disconnect components until the defective component is located that is shorting power a Disconnect Smart Summing Boxes and cycle power b Disconnect 12 a...

Страница 16: ...e of the displayable range 1 Verify programmed settings are correct a Verify Zero calibration is valid and in range b Verify load cell capacity L C Cap has been correctly programmed 2 Open Smart Summi...

Страница 17: ...t was intended for use as a Remote Display and not programmed incorrectly a Bin A LED should be solid in setup menu if unit is intended to be configured as remote display If unit is NOT intended to be...

Страница 18: ...accurate reading or proceed to Comprehensive Load Cell Inspection Procedure Comprehensive Load Cell Inspection Procedure 1 Record Note Current Weight Reading on BinTrac Indicator 2 Disconnect all but...

Страница 19: ...ets forth Buyer s sole and exclusive remedies for any defect in the goods The rights and obligation under this warranty may not be assigned or delegated to a third party by Buyer without the prior wri...

Страница 20: ...the declaration Models PS17 BT260 SSB100 The object of the declaration described above is in conformity with the relevant Union harmonisation legislation Applicable Directives Harmonized Standards or...

Отзывы: