MX1609/1609T/MX1609P thermocpuples modules
Up to 16 type K thermocouples (Ni-CrNi) can be connected to the module MX1609/MX1609P
for measuring temperatures.
Up to 16 type T thermocouples (Co
−
CoNi) can be connected to the module MX1609T for
measuring temperatures.
MX1609/MX1609T/MX1609P connectable transducers
Transducer type
Connector sockets
see page
Thermocouple type K, type T
1 ... 16
MX1609
Type
Thermocouple material 1 (+) Thermocouple material 2 (
−
)
K
Nickel-chrome (color code green) Nickel-aluminum (color code white)
+
−
Thermocouple
−
+
−
+
T
Copper (color code brown) Copper
−
Nickel (color code white)