110
Connection
HBM
QuantumX
6.10.1
MX1615 pin assignment
To enable insertion or removal of a transducer connection to be unmistakably identified, Pin
4 and Pin 5 in the connector plug must be bridged! This is automatically the case with all
bridge transducers; this bridge needs to be completed only in the event of voltage
measurements. If this bridge is missing, no measurement values will be recorded at the
connection!
1
2
3
4
5
6
7
8
1 2 3
1-Wire-EEPROM
(Maxim ) Bottom view
2 Data
3 no function
Bridge
1
Fig.6.15:
Pin assignment of the connector plug, view from the connection side
Pin
Connection
1
TEDS
2
Bridge excitation voltage (+)
3
Sense lead (+)
4
Bridge excitation voltage (
−
)
5
Sense lead (
−
)
6
Measuerment signal (+)
7
Measurement signal (
−
)
8
Housing (shield connection)