126
Connection
HBM
QuantumX
6.11.13
DC voltage sources 10 V and 60 V
Supported by modules: MX840, MX840A, MX410, MX440A
Voltage ranges supported by the amplifiers:
10 V and 60 V: MX840, MX840A, MX440A
10 V: MX410, MX1601, MX1615
Pin assignment for the MX1601 module see chapter 6.9.1
9
4
Hsg.
14
(
−
)
(+)
U
Cable shield
Supply voltage zero
Hsg.=Housing
6
1
1 2 3
1
5
6
10
11
15
2 Data
3 No function
1
1-wire EEPROM
(optional)
(view from below)
1
8
Push in
6
4
5
7
8
1
MX1615
1
8
Push in
1
5
2
8
4
MX1601
Maximum input voltage
to housing and signal
ground :
MX840, MX840A, MX440A:
"
60 V
MX1601:
"
40 V
Adjustable sensor supply:
5 V
…
24 V:
0 V:
Pin 12
Pin 11
Pin 6
Pin 7
MX840, MX840A und MX440A:
Two measuring ranges (10 V or 60 V) can be selected, depending on the parameterization.