107
Connection
HBM
QuantumX
6.9.1
MX1601pin assignment
So that insertion or removal of a transducer connection can be unmistakably identified, Pin 2
and Pin 5 in the connector plug must be bridged! If this bridge is missing, no measurement
values will be recorded at the connection!
1 2 3
1
2
3
4
5
6
7
8
1-Wire-EEPROM
(Maxim )
Bottom view
2 Data
3 no function
1
Bridge
Fig.6.13:
Pin assignment of the connector plug, view from the connection side
Pin
Connection
1
Voltage input 10 V (+), 100 mV (+), IEPE (+)
2
Signal ground, TEDS (
−
)
3
Current input 20 mA (+)
4
TEDS (+)
5
Always connect with Pin 2! (Plug identification)
6
Active sensor supply (+)
7
Active sensor supply (
−
)
8
Housing (shield connection)
NOTE
The transducer excitation voltage can be set in the range of 5 ... 24 V and is only
available at channels 1 ... 8. At channels 9 ... 16, the supply voltage (10 V ... 30 V) is
output less approx. 1 V. A current drain of max. 30 mA is possible; the current limiter
cuts the transducer excitation voltage at a higher current drain.
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...