154
Functions and outputs
HBM
QuantumX
Limit value monitoring
The limit value unit comprises eight limit value switches for monitoring of eight signals. Every
signal that is available throughout the system can be used as input signal.
The logic output can be assigned to a digital output. Parameterization includes the input
signal and, in addition, the switching threshold, hysteresis, switching logic and output logic.
The limit values’ update rate is 4800 Hz.
Switching on
Switching lebvel
Hysteresis value
Switching level
Switching on
Ausschalten
Switching off
Over limit
Below limit
Hysteresis value
24 V
0 V
24 V
0 V
LV1
Gr2
LV1 Logic 1
LV2 Logic 1
Fig. 7.5:
Limit value functions and parameters