81
Connection
HBM
QuantumX
Retrofitting the transducer plug with TEDS
The IEEE standard 1451.4 defines a generally accepted process which enables sensors to
be identified. The sensor is identified via the respective data sheet which is stored in
electronic format in the sensor, cable or plug on a 1
−
wire EEPROM (TEDS
−
Transducer
Electronic Data Sheet). The amplifier uses a serial 1
−
wire interface for communication with
this EEPROM, reads out the data sheet and configures the amplifier accordingly.
Retrofilling TEDS in the plug is shown in the figure below.
HBM empfiehlt den TEDS-Baustein (1-Wire
R
EEPROM) DS24B33 von Dallas Maxim. Um
diesen Chip lesen und beschreiben zu können ist mindestens Firmware-Version 1.21.16 /
2.21.16 sowie mindestens TEDS-Editor Version 3.3 nötig.
2 Data
3 No function
1
6
1
4
9
Bridge
1 2 3
1-wire EEPROM
(Maxim )
1
5
6
10
11
15
View from below
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...