20
Introduction
HBM
QuantumX
Synchronization using IRIG
−
B
IRIG
−
B is a standardized time code.
The digitally or analogously modulated time signal is externally fed to an arbitrary analog
voltage input of type MX840A or MX440A amplifiers to synchronize the QuantumX system
(see assignment, chapter 6.3.1).
The B127 format uses an analogons mudulation. The connection is similar to the one used
with a 10 V voltage sensor. The other formats are BCD-encoded and require connection
analogously to the sensor “Frequency measurement without directional signal” , see chapter
6.11.21.
The amplifiers can record type B000 through B007 and B120 through B127 IRIG
−
B signals.
All modules connected via FireWire are automatically synchronized as well. The code
includes the time, the year and, optionally, the seconds of the current day.
Comparison of synchronization mechanisms
Characteristic
FireWire
Ethernet (NTP)
EtherCat
IRIG
−
B
Synchronization with
other device types
QuantumX only
QuantumX, MGCplus
other
interrogators
all EtherCAT
nodes
all IRIG
−
B-
nodes
Max. distance
between QuantumX
modules
5 m (40 m with
FireWire extender,
500 m via fiber-optics
100 m (Ethernet)
100 m
−
Number of modules
to be synchronized
24
unlimited
CX27 required,
unlimited
unlimited
MX440A, MX840A
required
Synchronization
accuracy
< 1
μ
s
100
μ
s to 10 ms
< 1
μ
s
< 1
μ
s
Synchronization
setting time
immediately
approx. 2 h on first
start, approx. 10 min
on restart
immediately
immediately
Synchronization
Master
Auto
1 QuantumX module
external SyncMaster
, e.g. PC
external SyncMaster
external IRIG
−
B
master
Voltage supply
< 1.5 A, looped
through
−
−
−
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...