131
Connection
HBM
QuantumX
6.11.18
Thermocouples
Supported by modules: MX840, MX840A, MX440A, MX1609, MX1609
−
P
MX1609/MX1609P modules support only type K thermocouples; the MX1609
−
T module type
T supports type T with mini thermocouple plugs (see page 98). Cold
−
spot compensation is
integrated here.
With MX840, MX840A and MX440A, the accesory 1
−
THERMO
−
MXBOARD for cold spot
compensation needs to be integrated in the SubHD connector (see following page).
Type
Thermocouple material 1 (+)
Thermocouple material 2 (
−
)
J
Iron
Copper-Nickel
K
Nickel-chrome (color
−
code green)
Nickel-aluminum (color code white)
T
Copper
Copper-nickel
S
Rhodium-platinum (10%)
Platinum
E
Nickel-chrome
Copper-nickel
B
Rhodium-platinum (30%)
Rhodium-platinum (6%)
N
Nickel-chrome-silicone
1)
Nickel-silicone
R
Rhodium-platinum (13%)
Platinum
+
−
Thermocouple
1
−
THERMO
−
MX BOARD
(soldered into connector plug and contains a
cold spot compensation element and TEDS)
Maximum input voltage
to housing and signal
ground :
"
60 V
Connection with MX840, MX840A, MX440A
1
5
6
10
11
15
When recording temperatures with thermocouples using MX840, MX840A, MX440A, you
must use a connector with integrated cold spot compensation element
(1
−
THERMO
−
MX BOARD).
1)
Nicrosil
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...