8.4.1.2
ALARm:BATTery:ENABle?
Returns the state of the battery alarm enable.
Example: ALAR:BATT:ENAB?
Response: 0
A value of 1 is returned if the battery alarm is enabled. A value of 0 is returned
if the battery alarm is disabled.
8.4.1.3
ALARm:BATTery:ENABle <bool>
Sets the state of the battery alarm enable.
Example: ALAR:BATT:ENAB 1
The <bool> parameter turns the battery alarm enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.4.10, Pass-
word Commands).
8.4.1.4
ALARm:BEEP?
Returns the state of the audible alarm enable.
Example: ALAR:BEEP?
Response: 0
A value of 1 is returned if the alarm beep is enabled. A value of 0 is returned if
the alarm beep is disabled.
8.4.1.5
ALARm:BEEP <bool>
Sets the state of the audible alarm enable.
Example: ALAR:BEEP 1
The <bool> parameter turns the alarm beep enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.4.10, Pass-
word Commands).
8.4.1.6
ALARm:CLEar
Clears the alarm events.
Example: ALAR:CLE
The alarm clear command clears the alarm events.
8.4.1.7
ALARm:DATE:FIRSt?
Returns the date of the first alarm event.
86
Hart Scientific
8 Digital Communications Interface