8.4.2.12
CALCulate[<chn>]:PARameter<num>:RESolution <num>|MIN|MAX|DEF
Sets the display resolution.
Example: CALC:PAR2:RES 2
The PARameter suffix, <num>, specifies either temperature resolution (1) or
relative humidity resolution (2). The <num> parameter specifies the number of
decimal places to use. Specifying the MIN, MAX, or DEF parameter sets the
display resolution to the minimum, maximum, or default value respectively.
This command may be password protected (see Section 8.4.10, Password
Commands).
8.4.2.13
FETCh? [<chn>]
Returns a comma-delimited list of the most recent measurements for the speci-
fied channel or for both channels.
Example 1: FETC? 1
Response: 25.576,29.30
or
Response: 1,1,25.629,C,29.32,%,2003,9,16,11,2,38
Example 2: FETC?
Response: 25.582,29.32,26.341,37.96
or
Response: 1,1,25.629,C,29.29,%,2,0,C,0,%,2003,9,16,11,1,42
The <chn> parameter is optional and specifies the channel (1 or 2). If the
<chn> parameter is omitted, the most recent measurements on both channels
are returned. If a channel is disabled or no sensor is connected, the measure-
ment values returned for that channel are 0. This command does not affect the
measurement action.
If the extended format setting is OFF (see Section 8.4.2.15, FOR-
Mat:TDST:STAT), the measurement values are returned in a comma-delimited
format as follows:
<Ch1Temp>,<Ch1Humid>[,<Ch2Temp>,<Ch2Humid>]
The temperature values are always in the current temperature units. The humid-
ity values are always in %RH.
If the extended format setting is ON, the measurements are returned in a
comma-delimited format as follows:
100
Hart Scientific
8 Digital Communications Interface