7.2.1
Display Setting
The DISPLAY SETTING function allows you to select the temperature display
resolution (T RES), humidity display resolution (H RES), decimal preference,
unit of temperature and display contrast. These settings may be password-pro-
tected (see Section 7.5.4, Password).
The temperature and humidity display resolutions are selectable between 0, 0.1,
0.01, and 0.001. These settings do not afect the resolution of recorded or
printed data. Recorded and printed data are fixed at 0.01°C or °F and 0.1% RH.
The decimal preference is selectable between a period (.) and a comma (,).
The available units of measure are degrees C and degrees F.
The contrast can be adjusted here using the
lr
buttons. Setting the adjust-
ment number more negative lightens the screen and setting it more positive
darkens the screen. The contrast may also be adjusted from the main screen by
using the
ud
buttons (see Section 5.1, Front Panel).
The
lr
buttons are used to select the parameter setting. The
ud
buttons are
used to scroll between each parameter. Press the
Enter
button to change and
save the new settings. Press the
Exit
button to cancel any changes and exit to
the menu.
1620 the “DewK”
35
7 Menu Functions
Figure 14
Display Setting