Specifying the MIN, MAX, or DEF parameter returns the minimum, maxi-
mum, or default value respectively.
8.4.4.11
TRIGger:TIMer <num>|MIN|MAX|DEF
Sets the measurement period setting in seconds.
Example: TRIG:TIM 30
The <num> parameter can be one of the following values: 1, 2, 5, 10, 15, 30,
60, 120, 300, 600, 900, 1200, 1800 or 3600. If a value other than one listed
here is used, the instrument automatically selects the nearest acceptable value
that is less than the value entered. Values greater than 3600 or less than 1 are
ignored and cause a -222, “Data out of range” error. Specifying the MIN,
MAX, or DEF parameter sets the measurement interval timer to the minimum,
maximum, or default value respectively. The *RST command sets the measure-
ment period to 2 seconds. This command may be password protected (see Sec-
tion 8.4.10, Password Commands).
8.4.5
Channel Commands
The channel commands are used for switching channels on and off.
8.4.5.1
ROUTe:CLOSe? <chn>
Returns the state of the specified channel.
Example: ROUT:CLOS? 1
Response: 0
The <chn> parameter specifies the channel (1 or 2). A value of 1 is returned if
the specified channel is on. A value of 0 is returned if the specified channel is
off.
8.4.5.2
ROUTe:CLOSe <chn>
Switches on the specified channel.
Example: ROUT:CLOS 1
The <chn> parameter specifies the channel (1 or 2) to switch on. This com-
mand may be password protected (see Section 8.4.10, Password Commands)
8.4.5.3
ROUTe:OPEN? <chn>
Returns the state of the specified channel.
Example: ROUT:OPEN? 1
Response: 0