7.1
Channel Menu
The CHANNEL menu provides functions for selecting the channel, setting the
sensor identification, setting the sensor lock, and setting the sensor calibration.
The functions that appear in the CHANNEL menu are CHANNEL SETTING,
SENSOR ID, SENSOR LOCK, and SENSOR CAL.
7.1.1
Channel Setting
The CHANNEL SETTING function allows you to turn the channels on or off,
set the measurement period, turn averaging on or off, and set the time over
which the rate of change is calculated for measurements.
These settings may be password-protected to prevent them from being changed
(see Section 7.5.4, PASSWORD).
The channels are turned on or off by selecting the channel and then the ON or
OFF setting using the
lr
buttons.
MEA PER, measurement period, is the time interval at which the measurements
are updated on the display. You may select between 1, 2, 5, 10, 15, and 30 sec-
onds, 1, 2, 5, 10, 15, 20, and 30 minutes, or 1 hour for the measurement period.
AVERAGE may be turned to ON or OFF. When average is on, the temperature
and humidity values displayed are averaged over the measurement period.
When average is off, the displayed values are instantaneous measurements.
RATE T, rate time, determines the time interval over which the measurements
are used to calculate the rate of change, and is selectable between 30 seconds,
1, 2, 5, 10, 15, 20, 30, 45 minutes, or 1 hour. The temperature rate is displayed