2.
Allow four hours for settling after the temperature and humidity
set-points are reached before collecting data.
3.
Measure the temperature with the 2626 probe and compare it with the
temperature read from the reference thermometer in the chamber to ob-
tain the temperature measurement error:
Terror Tn
T
Tn
Treference Tn
(
)
(
)
(
)
=
−
2626
or
Measure the relative humidity using the 2626 probe and compare it
with the relative humidity read from the chamber to obtain the humid-
ity measurement error:
RHerror RHn
RH
RHn
RHchamber RHn
(
)
(
)
(
)
=
−
2626
9.5
Temperature and Humidity Adjustments
Adjustments to the 2626-x calibration parameters are required when measure-
ment errors are excessive or outside of the specification. The calibration pa-
rameters can be accessed using the SENSOR CAL function in the CHANNEL
menu. The password is required to change the parameters. Calculation of new
calibration parameters requires that the measurement errors first be obtained
according to the procedure above in section
Temperature and Humidity Mea-
surement Errors
. After the
Temperature and Humidity Adjustments
proce-
dure is completed, perform the
Temperature and Humidity Measurement
Errors
procedure again to verify the calibration.
1.
Calculate the adjustment
∆
TSL to the temperature slope parameter
TSL, and the new temperature slope parameter:
∆
TSL
Terror T
Terror T
T
T
= − ° ×
−
−
10
3
1
3
1
C
[
(
)
(
)]
(
)
TSL
TSL
TSL
′ =
+ ∆
2.
Calculate the slope-adjusted temperature errors:
Terror Tn
Terror Tn
TSL
Tn
(
)
(
)
(
)
′ =
+
×
− °
°
∆
25
10
C
C
3.
Calculate the median slope-adjusted temperature error, the adjustment
∆
TOS to the temperature offset parameter TOS, and the new tempera-
ture offset parameter:
1620 the “DewK”
141
9 Calibration Procedure