8.4.7.2
*OPT?
Returns a list of sensor models currently connected to the inputs.
Example: *OPT?
Response: “2626-S”,"0"
The model number of the sensor connected to input 1 is reported first. The
model number of the sensor connected to input 2 is reported second. If no sen-
sor is connected to an input, the model number is reported as “0". The sensor
model numbers are enclosed in quotes and separated by a comma.
8.4.7.3
*RST
Sets the instrument settings to defined conditions.
Example: *RST
The following parameters are set when this command is issued.
•
Sets the measurement period to 2 seconds.
•
Sets the measurement averaging on.
•
Sets printing to the serial port
•
Sets the temperature units to C.
•
Sets the temperature resolution to 2 decimal places and the relative hu-
midity resolution to 1 decimal place.
•
Stops auto logging.
•
Sets the statistical functions.
8.4.7.4
SYSTem:BOOT:VERSion?
Returns the boot ROM version.
Example: SYST:BOOT:VERS?
Response: 0.31
8.4.7.5
SYSTem:ERRor?
Returns a message from the system error queue.
Example: SYST:ERR?
Response: 0,"No error"
The error queue may contain up to ten messages. The messages are reported in
the order they occurred. Reading a message from the queue removes the mes-
1620 the “DewK”
115