The <chn> parameter specifies the channel (1 or 2). A value of 1 is returned if
the specified channel is off. A value of 0 is returned if the specified channel is
on.
8.4.5.4
ROUTe:OPEN <chn>
Switches off the specified channel.
Example: ROUT:OPEN 1
The <chn> parameter specifies the channel (1 or 2) to switch off. This com-
mand may be password protected (see Section 8.4.10, Password Commands)
8.4.6
Calibration Commands
The calibration commands are used to access the sensor calibration parameters.
The calibration commands are always password protected.
8.4.6.1
CALibrate<chn>:ALERt?
Returns the state of the sensor calibration expired alert enable for the specified
channel.
Example: CAL1:ALER?
Response: 0
The CALibrate suffix, <chn>, specifies the channel (1 or 2). A value of 1 is re-
turned if the calibration alert is enabled for the specified channel. A value of 0
is returned if the calibration alert is disabled for the specified channel.
8.4.6.2
CALibrate<chn>:ALERt <bool>
Sets the state of the sensor alert enable for the specified channel.
Example: CAL1:ALER 1
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <bool> pa-
rameter turns the sensor alert on (1 or ON) or off (0 or OFF). This command is
password protected (see Section 8.4.10, Password Commands).
8.4.6.3
CALibrate<chn>:DATE:CALibrate?
Returns the sensor calibration date for the specified channel.
Example: CAL1:DATE:CAL?
Response: 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is re-
turned in the format, <year>,<month>,<day>.