16
Measuring Bores Using the A-1511 Wand Bore Target/R-1308
Readout
There are two methods of obtaining readings when using the A-1511 Wand Bore Fixture. The first
method is by attaching the R-1308 Readout to the target. The second method is by using the R-1307
Readout for wireless communication. To configure the R-1307 Readout for dual-axis mode, see
Appendix B, beginning on Page 24.
Using a measuring tape, measure the diameter of the bore. Select the appropriate legs and/or extensions
that can be adjusted to the radius of the bore, less 3.750 in. (95.25 mm).
Attach these legs to the A-1511 Wand Bore Fixture.
Use a measuring tape to coarse-adjust the legs. The measurement of the legs to the center of the target
should be approximately the same as the radius of the bore. Final adjustments are made while measuring
the bore.