25
Configuring the R-1307 Readout for a Cabled (Local) Target
Note:
Shut
off power to the readout before connecting or disconnecting a target from the local port.
1.
Connect the cabled target to the local port of the readout
2.
Press and hold the MENU button for approximately 2 seconds to enter configuration mode.
3.
Set the Measurement Units
Press the MENU button until the upper display shows
Use the UP and DOWN arrow
keys to select either
for inches or
for millimeters.
4.
Set the Dampening Level
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow
keys to set the number of averages. Adjust this value as required to suit the application. The
default for this application is 8. For long distance shots, use 16 or 32.
5.
Set the Readout Function to Local Target
Press the MENU button until the upper display shows
. Use the UP and DOWN arrow
keys to select
.
6.
Select the PSD descriptor applicable to your target
Press the MENU button until the upper display shows
, where
designates the
target calibration factor number. There is one calibration record for each target purchased. The
R-1307 can store up to 9 records. Each calibration record in the R-1307 has the following target
types:
•
TGT=0 (for HLI use only. Do not use )
•
TGT = nn, P.10.10 (10x10 mm sensor- pulsed beam mode)
•
TGT = nn, F.10.10 (10x10 mm sensor- fixed beam mode)
•
TGT = nn, P.4.4 (4x4 mm sensor- pulsed beam mode)
•
TGT = nn, F.4.4 (4x4 mm sensor- fixed beam mode)
•
TGT = nn, P.20.20 (20x20 mm sensor- pulsed beam mode)
•
TGT = nn, F.20.20 (20x20 mm sensor- fixed beam mode)
•
TGT = nn, P.40.40 (40x40 mm sensor- pulsed beam mode)
•
TGT = nn, F.40.40 (40x40 mm sensor- fixed beam mode)
nn= R-1037 Readout number and matching target number
Press the UP or DOWN arrow to select the correct target number, which will change the second window.
For example,
or
for R-1307 #2. The PSD sensor size and type is
fixed to the target calibration record.
Warning: Targets are matched to specific calibration records in the R-1307 Readouts. For
example, Target #1 must be connected to Calibration Record #1 in the R-1307 or the calibration is
void. However, each R-1307 can have up to 9 target records, so up to 9 different target calibration
records can be stored in each R-1307. When there are multiple calibration records, the record ID
must match the target ID, so if you have Target #1, you should select TGT=01 to select the
matching calibration factors.
7.
To exit configuration mode, press and hold the MENU button for approximately three
seconds until the display returns to normal mode.
The R-1307 will also return to normal mode automatically after approximately four seconds of
inactivity.