19
Measuring Bores Using the A-1511 Wand Bore Target/R-1307
Readout
Ensure that the R-1307 Readout is configured correctly for the A-1519-2.4ZB Wireless Target. See
Appendix B beginning on Page 24 for configuration instructions.
Place the measuring tip on the side of the bore so it
is resting on the support leg. Adjust the
support
leg
until the horizontal reading on the R-1307 readout
is less than .030 in. (0.77mm). Adjust the
measuring
leg until the vertical reading is near
.000.
Slowly sweep the target in and out until the highest number displays on the R-1307 Readout. Press the
Zero Button
on the side of the target for approximately 3 seconds to zero the readout.