26
Configuring the R-1307 Readout for the A-1519-2.4ZB Wireless
Target in Dual-Axis Mode
1.
Press and hold the
Menu
button until the menu displays.
2.
Press and release the
Menu
button until
ID
displays in both the V and H Display windows.
3.
Use the up or down arrow keys to set the
Vertical ID
to 01.
4.
Press the
Menu
button again and use the up or down arrow keys to set the
Horizontal ID
to 01.
5.
Press the
Menu
button again and use the up or down arrow
keys to set the
Channel
to
6.
Note:
Ensure that the Target System ID dial on the side of the
target is set to 1.
7.
Press the
Menu
button until
displays in the Vertical Display window. Use the up or down
arrow keys until
displays in the Horizontal Display window. After a few seconds the R-1307
readout will start polling for the target. When the laser beam is on the target, readings display in the
Vertical and Horizontal Display windows.
8.
Press and hold the
Up
arrow key to zero the readout.
9.
Take measurements in each bore as described in
Measuring Bores with the A-1511 Wand Bore
Target and the R-1307 Readout
beginning on Page 19.
Note:
When using the Wand Bore Fixture with the R-1307 Readout, adjust the support leg until the horizontal
readings are less than .030 in. (.73 mm).
Содержание L-705