L-703SP - How It Works
Step-by-Step Surface Plate Calibration Procedure
Step 1 – Plate Setup
These parameters need to be entered:
•
the plate border size (default is 1.5 in. 37 mm).
•
the plate grade,
•
the plate size in X & Y,
•
the number of points to measure for the
horizontal, vertical and diagonal line segments,
•
pick the label location,
•
the plate serial number,
•
the temperature for the top and bottom of the
plate and,
•
the alignment tolerance.
Step 2 – Set Corner and Mid-Point Locating Tools
To make the process go faster, a corner-locating tool
is supplied to help align the straight edge to the line
segment.
Note
- there are slightly different tools for a 1.5-in.
border and a 2.5-in. border.
a.
The tool self-aligns to the plate’s corner and the
suction cup holds it in place.
b.
Next, use the locating tool to set the midpoint of
the 4 outside segments of the plate and mark the
point with a pencil.
c.
Place the midpoint location tool on the point for
use when you get to this segment
.