4
L-703S Level Vials
When using the L-703S in machine tools, we typically will do measurements with the laser
at the NORMal position (12:00) and then rotated with the spindle 180 degrees to the
INverted position (6:00). There are level vials on the
side
of the laser to help you know if
the laser is directly at 12:00 or 6:00. However, this is not typically used in surface plate
calibration, so they are not used in this application.
To use the level, rotate the laser 90 degrees so that the control panel is on the side (i.e. at
3:00 or 9:00). Then slowly rotate the spindle until the bubble is centered in the circle.
This indicates the laser is ready to be aligned to the rotation axis.
Figure 5 - L-703 Level Vial