8
Water Safety Kit (WSK)
8.2.3 Calibrating Temperature
ON/OFF
to switch the meter on.
Press
HOLD/ENT
Press and hold
CAL
for 3 seconds to
enter temperature calibration mode.
Press
CAL
to select
°C
or
°F
and the
upper display will flash.
Insert the probe into a solution of known
temperature and press
HOLD/ENT
to
adjust the reading.
After 5 seconds the calibration will be stored
and the meter will revert to normal mode.
8.2.4 Changing Batteries
Unscrew top casing to expose the
battery compartment.
Replace with 4
‘A76’
button cell
batteries, noting polarity.
Replace top casing.
8.2.5 Error Codes
•
Er.0
- temperature calibration is out
of range. Use solution from 0-50°C
•
Er.1
- pH calibration solutions are out
of range. Replace buffers with fresh. If
problem persists replace pH electrode
8.3 Pocket Conductivity Sensor
The
Pocket Conductivity Sensor
is a
waterproof digital instrument featuring dual
display of Conductivity and Temperature.
8.3.1 Before You Begin
Remove the electrode’s protective cap. Soak
the electrode for a few minutes in alcohol
to remove any oil stains on the electrodes.
Rinse thoroughly with de-ionised water
and shake to dry.
Press
ON/OFF
key to switch on the tester.
8.3.2 Range Selection
1
Switch off the tester. Press and hold
°C/°F
key and then switch on the tester
using
ON/OFF
key. Release
°C/°F
key.
2
The LCD shows the currently selected
Range (the default is
AUTO
) in the lower
display. The upper display shows the
maximum possible reading for the selected
range. Press
HOLD
key repeatedly until
you see the required range (PU, LO or HI).
8.3.3 Measurement
1
Press the
ON/OFF
key to switch on the
tester. The
‘MEAS’
indicator appears
when the tester is in measurement mode.
2
Dip the electrode into the test solution
making sure that it is fully immersed. Stir
to clear any trapped air bubbles from the
electrode and let the reading stabilise.
3
The upper display shows the main
reading (conductivity) of the solution,
automatically temperature compensated
(ATC) to normalised temperature of 25°C.
The lower display shows the temperature
of the solution.
8.3.4 Calibration
Auto Calibration
1
Make sure the tester is in measuring
mode. Press
INC
or
DEC
key to enter
conductivity calibration mode.
2 ‘CAL’
indicator appears in LCD. The display
briefly shows
‘CAL’
and the number of
points the tester will be calibrated.
14:30 Page 33
34