25
Set Date
To set the Date:
• Highlight
‘Set Date’
and press
‘OK’
• Use the
UP/DOWN
keys to set the day.
When correct press
‘OK’
• Use the
UP/DOWN
keys to set the
month. When correct press
‘OK’
• Use the
UP/DOWN
keys to set the year.
When correct press
‘OK’
Set Time
To set the Time:
• Highlight
‘Set Time’
and press
‘OK’
• Use the
UP/DOWN
keys to set the hour
(24 hour clock format). When correct
press
‘OK’
• Use the
UP/DOWN
keys to set the
minutes. When correct press
‘OK’
to
return to the Setup menu
Set Language
To set the Language:
• Highlight
‘Set Language’
and press
‘OK’
• Use the
UP/DOWN
keys to highlight
the required language: English, French,
Spanish and Mandarin
• When correct press
‘OK’
Check Calibration
Validate the temperature calibration of
the incubator for performance review.
The temperature is controlled by two
laser-trimmed thermistors which are
supplied factory calibrated and designed
for long term stability.
There is no natural drift in the calibration
of the incubator over time and hence no
need to field-adjust the calibration.
Using two identical thermistors located
separately provides a dual validation approach
to performance monitoring. It is possible
but unlikely that either thermistor may be
damaged in normal use but the possibility
of both thermistors being compromised in
an identical manner is almost zero.
The Check Calibration function provides a
simple electronic check that both thermistors
are reading the same temperature as each
other within a specified tolerance. If the
comparison is out of range the display will
show
Error 107: Validation
.
• Highlight
‘Check Calibration’
and press
‘OK’
• The incubator will display
‘Calibration
Verified’
upon completion
• Press
‘Back’
to return to the Setup menu
In the unlikely event of failure, return the
incubator to your local service representative
for attention, quoting the serial number
shown on the base.
Version
View the software version number for
identification and potential upgrade.
• Highlight
‘Version’
and press
‘OK’
• Note the software version displayed
Resuscitation Period
To improve microbiological analysis performance,
stressed samples should be subjected to a
period of acclimation prior to incubation.
To automatically include a resuscitation
period highlight
‘On’
and press
‘OK’
.
Resuscitation periods from 1-4 hours are set
from the Incubation menu when set to
‘On’
.
If resuscitation is not required highlight
‘Off’
and press
‘OK’
.
6
Incubation and Incubator Operation