6
Incubation and Incubator Operation
2
6.
.5 Download Incubation Data
Water quality reports can be enhanced with
downloaded data from the incubator detailing
temperature profile and power consumption.
To download data in a
Comma
Separated Values
(CSV) format:
• Connect the incubator via the
USB
cable
• Open the
Pota Pal app
• Select
‘Download Incubation Data’
• Choose
‘All Data’
to download all
available data or
‘Selected Data’
• Select the folder to store
the downloaded data
• For
‘Selected Data’
only, a list of
available data sets will be shown
Multiple data sets can be selected
using the shift or ctrl key.
• Once all required data sets have
been selected press
‘Download’
•
‘Download progress’
will be
shown on the dialogue box
• Upon completion the
‘Download
Finished’
message will appear
Downloaded
CSV files
can be opened with
any spreadsheet or word processing package
as a table of data for further manipulation.
6.
Troubleshooting
The Wagtech incubator monitors
performance during all stages of
performance and provides information
regarding unexpected condition as follows:
Error 102: Low Temp
The incubator has failed to reach the
temperature set point within the specified
time limit. Remove from areas of low
temperature/high cooling and accept the error.
Error 103: High Temp
The incubator temperature has exceeded the
set-point for a significant period. Remove
from direct sunlight or high temperature
conditions and accept the error.
Error 107: Validation
The internal validation of thermistor
temperature monitoring has exceeded the
tolerance of the device. Return to authorised
service centre for attention as soon as possible.
Error 110: Battery Low
The 12V DC battery has insufficient battery
capacity to complete the full incubation
cycle. Prepare alternative power supply
as soon as possible.
Error 111: Battery Critical
The 12V DC has approximately 10 minutes
battery life remaining. Find alternative
power source immediately.
Error 112: Power Loss
Unexpected power loss has interrupted
incubation. Check for cause if possible
and restart incubation when reliable
power is available.
A full list of error codes is found
in Appendix 2 - Troubleshooting.
Incubator Contamination
Should any hazardous material be spilt
onto or into the incubator, cleaning and
decontamination should only be carried
out with a damp cloth and mild detergent.
Disconnect power during cleaning and do
not invert the incubator.
Do not immerse the incubator under
any circumstance.
Do not use acetone or any abrasive/
aggressive/hazardous cleaning agents.
For advice regarding acceptable cleaning
agents contact
.
User Service
The Wagtech incubator contains no user
serviceable parts. The automated calibration
validation will confirm the incubator is
operating correctly.
For service please contact your local
Palintest representative or email to
.