17
6
Lock the membrane filter in place by
pushing the filter funnel firmly into
position in the blue rubber base. Take
care to not touch the internal surface
of the filter funnel with your hand.
7
Ensure the filter funnel is aligned correctly
in the
‘Filtration’
position, indicated by
the graphic on the side of the filter funnel.
8
Pour the water sample into the filter
funnel up to the 100ml graduation
(or less if using a smaller sample).
9
Discard the excess water from the sampling
cup then insert the filter funnel/base
assembly into position in the sampling cup.
Care is needed to prevent sample spillage
from the filter funnel when assembling.
NOTE:
If the sample is not to be collected
in the sampling cup part of the membrane
filtration unit, then the filter funnel/base
assembly can be inserted into position
in the sampling cup straight away.
10
Connect the hand vacuum pump
and silicone tubing to the MFU.
11
Use the pump to create a vacuum and
commence the filtration. The sample
level in the filter funnel will fall rapidly.
Do not pump too many times so as to
avoid drawing excess air through the
membrane filter. The water passes through
the pores in the membrane filter and
is collected in the sampling cup. Any
bacteria in the sample are collected on
the surface of the membrane filter.
5
Sample Processing - Membrane Filtration