Notice Messages and Troubleshooting
5-19
P/N 39551-00, Rev. F
395
SYSTEM PROBLEM
THE EFFLUENT TUBING IS NOT INSTALLED
The effluent tubing may not be installed.
1. Check tubing at Yellow valve.
2. Check tubing at White valve.
3. Check tubing at Green valve.
4. Check slide clamp at SPM.
5. Resume operation.
396
SYSTEM PROBLEM
THE PRODUCT BAG IS CLAMPED
The tubing to the product bag is clamped.
1. Check tubing at Green valve.
2. Check tubing for closed clamps.
3. Check tubing for kinks.
4. Resume operation.
397
BOWL TYPE MISMATCH
A different bowl size was detected. Confirm
whether the MCS+ device detected the correct
bowl:
1. Press YES to accept the new bowl size and
continue.
2. Press NO to keep the existing bowl size and
continue.
398
PROCESS PROBLEM
IS THE BUFFY BAG EMPTY?
Air has been detected at the BLAD,
indicating that the Buffy Bag may be empty.
1. Press YES to confirm and switch to donor.
or
2. Press NO to continue to draw from Buffy Bag.
399
PROCESS PROBLEM
IS THE BUFFY BAG EMPTY?
Air has been detected at the BLAD,
but the Buffy Bag should be empty.
1. Press YES to confirm and switch to donor.
or
2. Press NO to continue to draw from Buffy Bag.
400
SELF TEST PROBLEM
THE COMPUTER CPU TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
402
SELF TEST PROBLEM
THE COMPUTER ROM CRC TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
404
SELF TEST PROBLEM
THE PROTOCOL CARD CRC TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
405
SELF TEST PROBLEM
THE CALIBRATION CRC TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
406
SELF TEST PROBLEM
THE CONFIGURATION CRC TEST FAILED.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
#
Notice Messages
Help Messages