5-18
Notice Messages and Troubleshooting
P/N 39551-00, Rev. F
379
WEIGHER PROBLEM
EXPECTED FLUID FLOW TO WEIGHER NOT DETECTED.
No change in bag weight was detected.
1. Check that bags are properly installed.
2. Check tubing for occlusions.
3. Check tubing at valves.
4. Resume operation.
380
WEIGHER PROBLEM
EXPECTED FLUID FLOW FROM WEIGHER NOT DETECTED.
No change in bag weight was detected.
1. Check that bags are properly installed.
2. Check tubing for occlusions.
3. Check tubing at valves.
4. Resume operation.
381
MAXIMUM COLLECTION NOMOGRAM
TO BE EXCEEDED
Press RETURN to decrease the product volumes and remain
within the nomogram.
Press YES to collect the target product volumes, overriding the
nomogram.
The FDA Nomogram limit will be reached.
Decide whether to follow the nomogram and limit
the product volumes or to override the FDA
nomogram.
1. Press RETURN to follow the FDA nomogram.
2. Press YES to override the FDA nomogram.
387
WEIGHER PROBLEM
THE WEIGHT OF THE BAGS ON THE WEIGHER CHANGED.
An unexpected weight change occurred.
Do NOT touch the Weigher during operation.
1. Check that the bag hangs free.
2. Check tubing at valves.
3. Resume operation.
388
PROCEDURE RECOVERY
The POWER has been OFF for a short time.
A recovery sequence can be performed to
CONTINUE the last procedure if desired,
or a NEW procedure may be initiated.
389
PROCESS PROBLEM
RETURN CYCLE WAS NOT COMPLETED.
The bowl is not empty.
1. Press RETURN to complete return cycle.
2. Press STOP to stop draw cycle.
3. Press DRAW to start draw cycle.
390
PROCESS PROBLEM
INCOMPLETE CYCLE. NO PRODUCT COLLECTED.
No platelets were collected this cycle.
The cycle was interrupted early.
1. Press RETURN to continue.
393
ECV MAX
THE DONOR ECV LIMIT IS REACHED
The maximum allowed Extracorporeal Volume
(ECV) limit was reached.
A complete Return cycle must be performed.
1. Press RETURN to continue.
394
ECV LIMIT
xx% TBV Reached – Return Now?
The Extracorporeal Volume (ECV) limit was reached.
Decide whether to continue to Draw or to start
Return.
1. Press DRAW to keep drawing.
2. Press RETURN to go into return.
#