Notice Messages and Troubleshooting
5-5
P/N 39551-00, Rev. F
R
ECOVERY
P
ROCEDURE
If, for any reason, the system has been powered off or reset during a procedure,
the following screen will be displayed:
To perform a procedure recovery, the disposable can remain fully installed. The
MCS+ device will resume the current procedure after the YES key has been pressed.
At this time further information on how to perform a procedure recovery is available
in the on-line help screen. Full procedure statistics in the HaemoUpdate screen can
be retrieved by pressing the HELP key twice. Once the MCS+ device has performed
its self-tests, the MCS+ device can be guided through the centrifuge cover testing and
disposable installation. Upon successful completion of the self-tests and disposable
installation, the procedure may be resumed by pressing the RETURN key. The MCS+
device will now resume the procedure by returning the contents of the bowl and any
additional plasma to the donor or patient.
R
EPEAT
V
ENIPUNCTURE
When continued blood flow through the needle is no longer possible due to an
infiltration or any other cause, it may be desirable to return any red cells remain-
ing in the bowl. The following procedure allows for changing needles, perform-
ing a second venipuncture, returning any blood remaining in the disposable and
either continuing or discontinuing the collection procedure. The collection tar-
gets (Number of platelets, volume of plasma, etc.) for the procedure may not
have been met. Refer to the HaemoUpdate screen for current product informa-
tion.
PROCEDURE RECOVERY
The POWER has been OFF for a short time.
The POWER has been OFF for a short time.
Press YES for PROCEDURE RECOVERY.
Press NO to for NEW PROCEDURE.
Press HELP for More Information.
388
Содержание MCS+
Страница 2: ...P N 39551 00 Rev F...
Страница 10: ......
Страница 11: ...P N 39551 00 Rev F Chapter One INTRODUCTION OVERVIEW OF THE MCS DEVICE 1 2...
Страница 14: ......
Страница 44: ......
Страница 56: ......
Страница 86: ......
Страница 102: ......