System Description
2-9
P/N 39551-00, Rev. F
Modify
Using the MODIFY key, it is possible to change a protocol parameter before or
at any time during a procedure without interrupting the procedure.
To view or change the default parameters, press the MODIFY key. To select a pa-
rameter value, scroll through the list using the MODIFY key. Change a value us-
ing the + (YES) and – (NO) keys.
To permanently save the modified value(s), press the SAVE key. These values will
replace the default values. If you want to modify a value for only one procedure,
do not press SAVE. If your changes are not saved, they will be in effect only until
the system is powered off or the current procedure ends, at which time the system
reverts to the default parameters. Repeat this process each time you wish to mod-
ify the parameters.
Note:
Pressing the mode control key (DRAW, RETURN, or STOP) for the cur-
rent mode returns you to the main screen in the current mode without interrupt-
ing the procedure.
Operator
Configuration
Operator Configuration is available only in Platelet Protocol Software Version E
or later. Refer to the Platelet Protocol Manual, Part Number 51052-00 for details.
Procedure
Update
Modify
Parameters
Hints
SDPS
MODIFY PARAMETERS
Cycle 0
0 Platelet Volume
Ready To Go
0 Volume Processed
Cuff Pressure
Return Speed
50 mmHg
120 ml/min
Min. Plt. Volume
AC Ratio
200 ml
1: 9
Press STOP to return to the Main Display.
STOP
MODIFY selects, +/– to change, SAVE to save values.
AC Bag Volume
500 ml
NaCl Vol/Cycle
40 ml
✍