Maintenance
6-9
P/N 39551-00, Rev. F
Cleaning the
Blood, White,
and
Anticoagulant
Pumps
The pumps should be cleaned monthly or more frequently if needed. Always
clean the pumps after a fluid spill. A build-up of tubing debris or fluids may in-
terfere with pump efficiency. The following cleaning method is recommended.
1. Loosen the Pump Rotor screw located in the center of the Pump Roller
Assembly.
2. Lift the round Pump Rotor section out of the Pump Assembly.
3. Wash the entire Roller Assembly with cleaning solution and rub away any
debris on the Roller Assembly. Ensure that all the rollers turn freely on the
Roller Assembly.
4. Rinse with clear water to rinse away the cleaning solution.
5. Dry the assembly with a lint-free towel.
6. Replace the Roller Assembly into the Pump Well. It may be necessary to re-
move debris from the pumps between cleanings. A small brush has been pro-
vided in the tool kit for this purpose.
Cleaning the
Control Panel
The exterior of the MCS+ device should be cleaned with a mild detergent or dis-
infectant at regular intervals as well as any time a spill occurs. Do not use 100%
bleach directly on the machine; a 10% bleach/water solution or other disinfec-
tant may be used.
Figure 6-3, Pump Rotor Assembly
Pump Rotor
Screw