Notice Messages and Troubleshooting
5-11
P/N 39551-00, Rev. F
54
PUMP PROBLEM
WHITE PUMP POSITION IS NOT CORRECT.
A restriction may have occurred.
1. Check pump for restrictions.
2. Check pump for kinked tubing.
3. Resume operation.
56
PUMP PROBLEM
THE AC/BLOOD PUMP RATIO IS INCORRECT.
An occlusion may have occurred at the
Blood Pump or AC Pump.
1. Check both pumps for occlusions.
2. Resume operation.
57
PUMP PROBLEM
THE AC/BLOOD PUMP RATIO IS INCORRECT.
An occlusion may have occurred at the
Blood Pump or AC Pump.
1. Check both pumps for occlusions.
2. Resume operation.
58
VALVE PROBLEM
THE RED VALVE WAS DETECTED OUT OF POSITION.
1. Toggle the valve manually.
2. Resume operation.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
61
SYSTEM PROBLEM
THE COMPUTER HAS ENCOUNTERED AN ERROR.
A CPU error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
62
ECV LIMIT
THE DONOR ECV LIMIT IS REACHED.
The allowed Extracorporeal Volume (ECV)
limit was reached.
A complete Return cycle must be performed.
1. Press RETURN to continue.
63
SYSTEM PROBLEM
THE SAFETY SYSTEM HAS ENCOUNTERED AN ERROR.
1. Turn power OFF, then ON and try again.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
64
SYSTEM PROBLEM
THE SAFETY SYSTEM HAS ENCOUNTERED AN ERROR.
A Watchdog error caused the MCS+ device to reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
65
SYSTEM PROBLEM
THE SAFETY SYSTEM HAS ENCOUNTERED AN ERROR.
A DPM/SPM connector caused the MCS+ device to
reset.
1. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
66
SYSTEM PROBLEM
THE SAFETY SYSTEM HAS ENCOUNTERED AN ERROR.
A pressure error caused the MCS+ device to reset.
1. Check installation of DPM and SPM.
2. Power off and complete a
Procedure Recovery.
NOTE: If problem persists,
call Haemonetics Technical Hot Line.
#
Notice Messages