63
ModBUS Register Information
Setup/Relay 2
Type
40093
Unsigned
Integer
1
R/W
Selects the relay type (0 = Alarm; 1 =
Control; 2 = Status; 3 = Timer; 4 = Event)
Setup/Relay 2
Transfer
Setting
40094
Unsigned
Integer
1
R/W
Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)
Setup/Relay 2/Alarm
High Alarm
40095
Float
2
R/W
Sets the high alarm setpoint
Setup/Relay 2/Alarm
Low Alarm
40097
Float
2
R/W
Sets the low alarm setpoint
Setup/Relay 2/Alarm
High
Deadband
40099
Float
2
R/W
Sets the high alarm deadband
Setup/Relay 2/Alarm
Low
Deadband
40101
Float
2
R/W
Sets the low alarm deadband
Setup/Relay 2/Alarm
On Delay
40103
Unsigned
Integer
1
R/W
Sets the on delay time
Setup/Relay 2/Alarm
Off Delay
40104
Unsigned
Integer
1
R/W
Sets the off delay time
Setup/Relay 2/Control
Setpoint
40105
Float
2
R/W
Sets the controller setpoint
Setup/Relay 2/Control
Phase
40107
Unsigned
Integer
1
R/W
Sets the controller phase
(0 = Low; 1 = High)
Setup/Relay 2/Control
Deadband
40108
Float
2
R/W
Sets the controller deadband
Setup/Relay 2/Control
Overfeed
Timer
40110
Unsigned
Integer
1
R/W
Sets the overfeed timer value (mins)
Setup/Relay 2/Control
On Delay
40111
Unsigned
Integer
1
R/W
Sets the on delay time (sec)
Setup/Relay 2/Control
Off Delay
40112
Unsigned
Integer
1
R/W
Sets the off delay time (sec)
Setup/Relay 2/Control
Reset
Overfeed
Timer
40113
Unsigned
Integer
1
R/W
Resets the overfeed timer
Setup/Relay 2/Event
Setpoint
40114
Float
2
R/W
Sets the event setpoint
Setup/Relay 2/Event
Phase
40116
Unsigned
Integer
1
R/W
Sets the event phase (0 = Low; 1 = High)
Setup/Relay 2/Event
Deadband
40117
Float
2
R/W
Sets the event deadband
Setup/Relay 2/Event
On Max Time
40119
Unsigned
Integer
1
R/W
Sets the max on time (mins)
Setup/Relay 2/Event
On Min Time
40120
Unsigned
Integer
1
R/W
Sets the min on time (mins)
Setup/Relay 2/Event
Off Max Time
40121
Unsigned
Integer
1
R/W
Sets the max off time (mins)
Setup/Relay 2/Event
Off Min Time
40122
Unsigned
Integer
1
R/W
Sets the min off time (mins)
Setup/Relay 2/Timer
Hold Type
40123
Unsigned
Integer
1
R/W
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Setup/Relay 2/Timer
Sensor Select
40124
Unsigned
Integer
1
R/W
Selects which Sensor outputs are being
held/transferred during the timers on time
(this is used when Hold type is set for
single Sensor)
Setup/Relay 2/Timer
Hold Mode
40125
Unsigned
Integer
1
R/W
Selects hold outputs vs. set transfer value
during timers on time
Table A-8 Controller ModBUS Registers (continued)
Group Name
Tag Name
Register # Data Type
Length
R/W
Description