Operation
37
5.5 Verify using dry standard CVM
Select
Menu level
Accept
1.
MAIN MENU
2.
SENSOR-SETUP
3.
SELECT SENSOR
(if there is more than one sensor)
4.
VERIFY
5.
STANDARD
6.
OUTPUT MODE information
7.
TURN OFF SAMPLE INLET
8.
DRAIN MEAS.CHAMBER
(via the lower feed union, then dry carefully)
9.
INSERT
CVM STANDARD
(For this purpose open measuring chamber,
remove sealing plug and wiper holder, dry
measuring chamber.
Insert the CVM module into the measuring
chamber. The arrow mark must point
downwards!
Ensure that the spring-loaded ball engages by
turning slightly in the sample feed opening on
the measuring chamber, see Fig.)
10.
x.xxx TRBFNU
11.
REMOVE CVM MODULE
12.
REPLACE WIPER
HOLDER AND CAP
13.
TIGHTEN CAP
14.
CONNECT SAMPLE INLET
15.
VERIFIC. READY
OPEN INLET
16.
OUTPUT
ACTIVE