Installation
23
3.5 Connecting sensor cable
You can connect the sensor cable to the controller very easily using the plug.
Retain the protective cap for the socket in case you need to remove the sensor in
the future. Connecting cables are available in the lengths 5 m, 10 m, 15 m, 20 m,
30 m and 50 m (see list of spare parts). From a length of 100 m a termination box
must be integrated, see Fig. 3-23.
Fig. 3-20
Connection of the sensor plug to the controller
You must note the following for the permanent connection of the sensor:
1.
Open the front flap.
2.
Remove the terminal strip J5 and remove the related cable from this terminal
strip.
3.
Undo the nuts for the sensor sockets on the inside of the housing and remove
the sensor sockets and cables from the housing (see Fig. 3-21).
Fig. 3-21
Removing sensor socket