40
Operation
5.7.1 Holding outputs / substitute values
The analogue current outputs can also retain the last measured values during
normal measurement operation. Proceed as follows to hold the current outputs
until another command is issued:
During the calibration the analogue current outputs can reflect the calibration
measurements, hold the last value or output a substitute value (Sec. 5.7, 5.7.1).
5.7.1.1 Enabling outputs
Select
Menu level
Accept
1.
MAIN MENU
2. TEST/MAINT
3.
HOLD OUTPUTS
4.
OUTPUT MODE
5.
HOLD OUPUTS or XFER OUTPUTS
6:
SET CHANNELS
7.
Choose sensor
8.
ACTIVATION
9.
START
Select
Menu level
Accept
1.
MAIN MENU
2. TEST/MAINT
3.
HOLD OUTPUTS
4.
ACTIVATION
5.
RELEASE