65
ModBUS Register Information
Setup/Relay 3/Event
On Min Time
40160
Unsigned
Integer
1
R/W
Sets the min on time (mins)
Setup/Relay 3/Event
Off Max Time
40161
Unsigned
Integer
1
R/W
Sets the max off time (mins)
Setup/Relay 3/Event
Off Min Time
40162
Unsigned
Integer
1
R/W
Sets the min off time (mins)
Setup/Relay 3/Timer
Hold Type
40163
Unsigned
Integer
1
R/W
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Setup/Relay 3/Timer
Sensor Select
40164
Unsigned
Integer
1
R/W
Selects which Sensor outputs are being
held/transferred during the timers on time
(this is used when Hold type is set for
single Sensor)
Setup/Relay 3/Timer
Hold Mode
40165
Unsigned
Integer
1
R/W
Selects hold outputs vs. set transfer value
during timers on time
Setup/Relay 3/Timer
Duration Time
40166
Unsigned
Integer
1
R/W
Sets the timer on duration time (sec)
Setup/Relay 3/Timer
Period Time
40167
Unsigned
Integer
1
R/W
Sets the period between timer on events
(mins)
Setup/Relay 3/Timer
Off Delay
40168
Unsigned
Integer
1
R/W
Sets the time the affected Sensor outputs
are held/transferred after the timer turns
off (sec)
Setup/Relay 3/Status
Level
40169
Unsigned
Integer
1
R/W
Sets the status level which will trigger
the relay
Comm/Net Card
Mode
40170
Unsigned
Integer
1
R/W
Sets the Modbus mode
(0 = RTU; 1 = ASCII)
Comm/Net Card
Baud
40171
Unsigned
Integer
1
R/W
Sets the Modbus baud rate (0 = 9600;
1 = 19200; 2 = 38400; 3 = 57600;
4 = 115200)
Comm/Net Card
Stop Bits
40172
Unsigned
Integer
1
R/W
Sets the number of stop bits (1,2)
Comm/Net Card
Data Order
40173
Unsigned
Integer
1
R/W
Sets the register data order for floats
(0 = Normal; 1 = Reversed)
Comm/Net Card
Min
Response
Time
40174
Unsigned
Integer
1
R/W
Sets the minimum response time
(0 to 30 sec)
Comm/Net Card
Max
Response
Time
40175
Unsigned
Integer
1
R/W
Sets the maximum response time
(100 to 1000 sec)
Comm/Net
Card/Addresses
sc100
40176
Unsigned
Integer
1
R/W
Sets the sc100 Modbus Address
Comm/Net
Card/Addresses
Sensor 1
40177
Unsigned
Integer
1
R/W
Sets the Sensor 1 Modbus Address
Comm/Net
Card/Addresses
Sensor 2
40178
Unsigned
Integer
1
R/W
Sets the Sensor 2 Modbus Address
Comm/Net Card/Stats
Good
Messages
40179
Unsigned
Integer
2
R/W
Number of good messages
Comm/Net Card/Stats
Bad
Messages
40181
Unsigned
Integer
2
R/W
Number of failed messages
Comm/Net Card/Stats
% Good Mesg
40183
Float
2
R/W
% of good messages
Table A-8 Controller ModBUS Registers (continued)
Group Name
Tag Name
Register # Data Type
Length
R/W
Description