50
Section 6
Maintenance
It has been possible to reduce the scope of the maintenance to be performed by
the user to only a few actions. Clearly summarised in a table and described in
detail in the following sections, these actions can be quickly and straightforwardly
performed by qualified personnel.
6.1 Maintenance schedule
6.2 Cleaning measuring chamber
The cleanliness of the measuring chamber is crucial for correct measured results.
This cylindrical optical measuring chamber has a rotating wiper with three wiper
profiles (only on plus version); these prevent normal soiling of the optical system
as it starts to build up. With very stubborn deposits, manual cleaning with a
suitable cleaning agent (e.g. citric acid) is recommended.
Note: Observe safety regulations and wear safety clothing!
•
Safety glasses
•
Gloves
•
Overall
Maintenance task
Interval
Clean measuring chamber:
ULTRATURB sc
ULTRATURB plus sc
Dependent on the substances contained
in the water
Replace wiper profile
(only plus version)
after 1200 cycles
Replace desiccant
Every 2 years
Monitor test equipment (CVM dry
standard)
Every 2 years
(factory test with certification)
Check zero point
Dependent on the substances contained
in the water
Check gradient
At least once a year
Select
Menu level
Accept
1.
MAIN MENU
2.
SENSOR-SETUP
3.
SELECT SENSOR
(if there is more than one sensor)
4.
TEST/MAINT
5.
MAINT.PROC.
6.
CLEAN MEAS.CHAMBER