84
Expanded programs
6.1.1.3
Calibration settings for single and Multi Wavelength mode
A method is calibrated by determining the absorbance values of
several standard solutions of known concentration.
Create and save a calibration curve by entering
values
or
measuring standards
or
entering formulas
(
Table 9
).
Calibration by "Enter values"
1.
Select
Enter values
.
2.
Press
Continue
.
3.
Press
+
.
4.
Enter the standard concentration.
5.
Press
OK
to confirm.
6.
Enter the corresponding absorbance value.
7.
Press
OK
to confirm.
8.
If necessary, repeat until all values have been entered.
Note: To change a value in the table, mark the corresponding row. Press
the key of the unit (e. g. mg/L) or Abs. Enter the changed value.
Table 9 Calibration settings
Option
Description
Enter values
Create a calibration table by entering the concentration values and the corresponding
absorbance values of the standard solutions be measured. The absorbance values are applied
in a diagram against the concentrations of the standard solutions. The calibration curve is
graphically presented (page
84
).
Measure standards
Create a calibration table by entering the concentration values of the standard solutions to be
measured and then determining the corresponding absorbance values by measuring the
standard solutions. The absorbance values are applied in a diagram against the concentrations
of the standard solutions. The calibration curve is graphically presented (page
85
).
Enter formula
If it is possible to determine the calibration curve from the mathematical relationship between
concentration and absorbance through linear regression or another curve adaptation, the
respective formula (linear, polynomial 2nd or 3rd order can be selected from a list and the
corresponding factors can be entered (page
86
).
Содержание DR 6000
Страница 2: ......
Страница 16: ...16 General information ...
Страница 132: ...132 Expanded programs ...
Страница 138: ...138 Maintenance 7 4 Replace fuses 1 Remove the cover below the power switch containing the fuses ...
Страница 142: ...142 Troubleshooting ...
Страница 145: ...145 Section 10 Contact ...
Страница 146: ...146 Contact ...
Страница 152: ...152 Help Guide ...
Страница 156: ...156 Index ...