![Hach DR 6000 Скачать руководство пользователя страница 141](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875141.webp)
141
Troubleshooting
File for instrument update is faulty.
Error during update.
Save the update file again and repeat
the procedure.
Error copying from USB memory
stick.
Error during update
Start the procedure again or contact
technical support.
No instrument backup present!
Check the USB memory stick.
Insufficient memory for update .
Error during update.
Select a memory with more space.
Update file is faulty.
Error during update.
Save the update file again and repeat
the procedure.
USB memory stick is not connected.
Update not possible.
Check the USB memory stick.
Network connection errors
Please check network configuration.
Please check the connection.
Please check the connection and
contact the administrator.
Network setup or FTP error
Error when calling up the local IP
address.
Network setup: DHCP client has no
connection to the DHCP server
Enter the IP address again.
Error during default gateway setup.
Network setup: default gateway cannot
be set for fixed IP address
Try to create the connection again.
Error during network drive setup!
Error during network setup
The target directory does not exist.
Check the settings.
Define target directory.
Error during subnet mask setup.
Network setup: Subnet mask cannot be
set for fixed IP address
Enter the subnet mask again.
Error in FTP connection.
FTP error
Make sure that the instrument is
connected to the network.
Network switched off.
Network setup off, when access to Viper
homepage via sidebar
Activate the online connection.
Remote server cannot be reached.
Error during network setup
Make sure that the instrument is
connected to the network.
The fixed address of the DR 6000 is not
accepted.
Remote server unreachable!
The fixed address of the DR 6000 is not
accepted.
The server name entered is incorrect.
Switch to "Automatic".
Enter the correct server name.
Web server cannot be reached.
Viper homepage cannot be reached.
Try the connection again later.
Error displayed
Cause
Resolution
Содержание DR 6000
Страница 2: ......
Страница 16: ...16 General information ...
Страница 132: ...132 Expanded programs ...
Страница 138: ...138 Maintenance 7 4 Replace fuses 1 Remove the cover below the power switch containing the fuses ...
Страница 142: ...142 Troubleshooting ...
Страница 145: ...145 Section 10 Contact ...
Страница 146: ...146 Contact ...
Страница 152: ...152 Help Guide ...
Страница 156: ...156 Index ...