Standard programs
3.
Press
Web Function
on the toolbar.
4.
Press
Instrument Update
.
The update will start automatically.
Note: When the instrument software has been updated, a prompt to
restart the instrument is displayed.
5.5.5.5
Program a new test
1.
The instrument needs to be at the "Main Menu" or press
Barcode Programs
in the Main Menu.
2.
Insert the zero cell or sample cell (corresponding to the
process specification) into the cell compartment.
The message "Program not available" is shown, followed by
a data overview with the specifications of the test that is to be
revised.
The process specification of the corresponding test contains
the new data (Wavelength, Factors, Reading Ranges,
Conversion Factor for chemical form etc.).
3.
Highlight the corresponding line including the specifications
of the test that is to be programmed and press
Edit
.
4.
To finish programming, press
Store
and then
Cancel
to
return to the Main Menu.
5.
Press
Start
to begin the test with the new data.
Note: Once the instrument has an active network connection, new
software versions are announced as they become available. An
instrument update can then be performed.
Содержание DR 6000
Страница 2: ......
Страница 16: ...16 General information ...
Страница 132: ...132 Expanded programs ...
Страница 138: ...138 Maintenance 7 4 Replace fuses 1 Remove the cover below the power switch containing the fuses ...
Страница 142: ...142 Troubleshooting ...
Страница 145: ...145 Section 10 Contact ...
Страница 146: ...146 Contact ...
Страница 152: ...152 Help Guide ...
Страница 156: ...156 Index ...