![Hach DR 6000 Скачать руководство пользователя страница 120](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875120.webp)
120
Expanded programs
7.
Place the analysis cell in the cell compartment.
8.
Close the cell compartment cover.
9.
Press
Read
.
10.
Remove the analysis cell.
11.
Close the cell compartment cover.
12.
Press
Zero
.
13.
Place the analysis cell in the cell compartment.
14.
Close the cell compartment cover.
15.
Press
Read
.
16.
Repeat this procedure until 5 comparison measurements
have been obtained.
The result will be displayed.
17.
Compare the result with the analysis cell required values.
18.
Press
Cancel
to return to Optical Checks.
6.7.3.5
Noise Check
With the noise check the photometric noise in the device is
checked.
This test can be used to test the noise at any wavelength during
use of a sample that corresponds to the absorbance level of this
wavelength.
1.
Press
Noise Check
in the "Optical Checks" menu.
2.
Press
to select the wavelength.
3.
Enter the wavelength and press
OK
to confirm your
selection.