118
Expanded programs
10.
Follow the directions on the display and set various filters
one after another in the order displayed.
11.
Press
Next
after inserting a filter.
After the last reading the results are displayed.
12.
Press
PC & Printer
icon
to send the data to a USB memory
stick, PC or to a printer.
The files will be stored automatically as CSV files (Comma
Separated Value). The file name will be formatted as
"Verification.csv".
6.7.3.2
Wavelength check
The Wavelength Check test is used to check wavelength
accuracy at 656.1 nm and 486.0 nm. The bandwidth is checked
at 656.1 nm.
1.
Select
Wavelength Check
from the "Optical Checks" menu.
2.
Press
to select the wavelength.
3.
Press
OK
to confirm.
4.
Press
Start
.
Note: During the UV lamp warm-up phase, the text "Lamp
Warmup"is shown.
The result is shown.
5.
Press
Cancel
to go back to the Optical Checks.