Geophysical Survey Systems, Inc.
SIR® 4000
Manual
MN72-433 Rev F
149
MLF 360 cm
Low Frequency antenna length 3.6m
Data Collection Mode: Point
Range: 750ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 10
Vertical Low Pass Filter: 60
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 12 KHz
MLF 480 cm
Low Frequency antenna set to a length of 4.8m
Data Collection Mode: Point
Range: 1000ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 6
Vertical Low Pass Filter: 40
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 12KHz
MLF 600 cm
Low Frequency antenna set to a length of 6.0m
Data Collection Mode: Point
Range: 1000ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 1
Vertical Low Pass Filter: 50
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 12 KHz
Borehole 120 MHz
Borehole antenna frequency 120 MHz. Note:
The borehole antennas are unshielded.
Data Collection Mode: Point
Range: 500ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 30 MHz
Vertical Low Pass Filter: 240 MHz
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 64 KHz
Borehole 300 MHz
Borehole antenna frequency 300 MHz.
Data Collection Mode: Point
Range: 300ns
Samples per Scan: 512
Resolution: 32 bits
Number of gain points: 5
Vertical High Pass Filter: 38 MHz
Vertical Low Pass Filter: 600 MHz
Scans per second: 32
Stacking: 32 scans
Transmit Rate: 64 KHz
Содержание SIR 4000
Страница 1: ......
Страница 4: ......
Страница 116: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 110 ...
Страница 140: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 134 ...
Страница 150: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 144 ...
Страница 156: ...Geophysical Survey Systems Inc SIR 4000 Manual MN72 433 Rev F 150 ...