Settings
The Settings menu option may be used to define the parameters used to collect
the tympanogram. A Settings dialog is presented with the tympanogram
parameters that are available to be selected. The tympanogram default settings
may be defined in the PC Config App. The device default settings may be found
in the Appendix.
Start Pressure
The Start Pressure Test Options button is used to select the starting pressure
(daPa) for collecting a tympanogram. When the button is selected, a dropdown
list of starting pressures is presented.
Range
The Pressure Range Test Options button controls the minimum and maximum
pressure that is displayed on the tympanogram. The button label indicates the
current setting. The button toggles between the options of Normal and Wide.
The Normal range is from +200 daPa to -400 daPa. The Wide range is from
+400 daPa to -600 daPa.