background image

GSI Instrument Services 

 

Description

 

The  GSI  Instrument  Services  allows  electronic  transmission  of  test  parameter 
information from the TympStar Pro to an external computer with a single push of 
the Data Transfer button. See the GSI Instrument Services user manual for detail 
on how to utilize its functionality. 
 

Operation

 

Data  capture  occurs  when  the  Store  pushbutton  is  pressed.  When  there  are  test 
results, comments or patient demographics saved in the TympStar Pro, data may 
be electronically transferred to a software solution on an external computer using 
the Data Transfer button.  
 

Public Interface (Direct)

 

The  Public  Interface  option,  provided  through  the  GSI  Instrument  Service, 
transfers the audiometric data from the  TympStar Pro in an XML format which 
may be directly incorporated into  an Electronic Medical Record. The GSI Suite 
utilizes this format. Alternatively, independent software programming engineers 
may implement the XML schema provided by GSI into their proprietary software 
in  order  to  manage  patient  data  directly.  The  direct  transfer  of  data  gives  the 
physician  immediate  access  to  the  data  in  the  electronic  record.    More 
information may be found on the Instrument Services USB that was included in 
the original shipment of the TympStar Pro or contact your GSI representative.  
 

Data Port (Direct)

 

The Data Port provides backwards compatibility with the TympStar (serial) data 
stream. This will require the selection of an available COM port. Using the Data 
Port interface makes it possible to transfer audiometric data from the TympStar 
Pro  directly  into  existing  Electronic  Medical  Record  solutions.  Independent 
software  programming  engineers  may  implement  the  data  stream  protocol 
provided  by  GSI  into their  proprietary  software  in  order to  manage  patient  data 
directly.  The direct transfer of data gives the physician immediate access to the 
data in the electronic record. More information may be found on the Instrument 
Services CD that was included in the original shipment of the  TympStar Pro or 
contact your GSI representative. 
 
 

 
GSI Suite 

GSI  Suite  Audiometric  Data  Management  software  (Rev.  2.3  and  higher)  is 
compatible with the TympStar Pro as well as legacy products. GSI Suite imports, 
saves,  and  stores  data  from  the  TympStar  Pro  and  allows  the  addition  of 
comments into a report.  The report data is saved in a PDF  or other format that 
may  be  saved  to  the  local  PC,  a  remote  location  or  attached  with  electronic 
medical  data  records  (EMR).  GSI  Suite  may  be  used  as  a  stand-alone  software 
solution or in combination with Noah 4 or OtoAccess.  
 

 

 

Содержание TYMPSTAR PRO

Страница 1: ...andard www grason stadler com Grason Stadler 10395 West 70th Street Eden Prairie MN USA 55344 800 700 2282 952 278 4402 fax 952 278 4401 e mail info grason stadler com GSITYMPSTAR PRO USER MANUAL Part...

Страница 2: ...D 0105594 Rev G i...

Страница 3: ...tion in this publication is proprietary to Grason Stadler This product incorporates software developed by others and redistributed according to license agreements Copyright notices and licenses for th...

Страница 4: ...wnout and Interruptions 7 Warning Connections 7 Warning Battery Safety 7 Cautions General 7 Shutdown Procedure 7 Recycling Disposal 8 Customer Responsibility 8 Warranty 9 Chapter 1 Introduction 10 Ind...

Страница 5: ...top 24 Start 24 Pressure Control 24 Reflex Stimulus Controls 25 Stimulus 25 Intensity 25 Present 25 Print 25 Data Transfer 27 Configure 27 Chapter 5 Test Type Displays 29 Display Monitor 29 Test Type...

Страница 6: ...s 103 Calibration Quick Check for Probe 103 Biological Calibration Check 105 Preventive Maintenance 105 Cleaning the System 105 Cleaning and Disinfecting Agents 106 Cleaning patient contact devices 10...

Страница 7: ......

Страница 8: ...G label identifies conditions or practices that may present danger to the patient and or user CAUTION The CAUTION label identifies conditions or practices that could result in damage to the equipment...

Страница 9: ...of Manufacture Attention consult accompanying documents China RoHS symbol for products with a 50 year life cycle B Patient Applied Part according to IEC 60601 1 On Off Next to power mains Keep Dry Thi...

Страница 10: ...directions for use A copy of the operating manual is available on this website www grason stadler com A printed copy of the operating instructions can be ordered from Grason Stadler for shipment withi...

Страница 11: ...he direct supervision by the aforementioned specialist Users should use their professional skills when interpreting the results and this should be done in conjunction with other testing as deemed appr...

Страница 12: ...ts in excess of those allowed by the standard and present a potential electrical shock hazard to the person being tested This icon indicates that the TympStar Pro is in compliance with Class 1 Type B...

Страница 13: ...amage to the power unit Follow all safety standards set by each place of employment NOTE If the instrument is connected to a PC power to the monitor and computer must be controlled by the isolation tr...

Страница 14: ...setup instructions Switch off the system power before connecting or disconnecting any system component s or accessories Warning Battery Safety This instrument contains a coin type lithium battery for...

Страница 15: ...may be broken or missing or are visibly worn distorted or contaminated should be replaced immediately with clean genuine replacement parts manufactured by or available from GSI This product should no...

Страница 16: ...ler facility If field service is requested there will be no charge for labor or material however there will be a charge for travel expense at the service center s current rate NOTE Changes in the prod...

Страница 17: ...plugged at the entrance or if fluid is running from the ear canal tympanometry should not be attempted until the condition is cleared Testing should not be performed on patients with conditions liste...

Страница 18: ...s subjected to varying degrees of stress which alters mobility of the eardrum Maximum mobility will occur when the pressure on both sides of the eardrum are equal Changes in mobility of the eardrum te...

Страница 19: ...Latency Test Multiple Frequency Tympanometry 250 Hz to 2000 Hz Operators have a choice of using GSI preprogrammed test parameters or programming customized test criteria A large touch enabled liquid...

Страница 20: ...iginal container if possible and return it prepaid to the factory for the necessary adjustments NOTE Keep the original packing material and shipping container so the instrument can be well packaged if...

Страница 21: ...he factory default test mode to set user defined power up setting see the information regarding the PC configuration software and the probe green lamp will begin to blink indicating that the instrumen...

Страница 22: ...scription Graphic B1 Mains Power Input IEC 14 WARNING To avoid the risk of electric shock this equipment must only be connected to supply mains with protective earth B2 USB Computer Connection USB B s...

Страница 23: ...resolution in order to maintain the aspect ratio of the tympanogram B6 Patient Response Inputs 6 35 mm mono jack for hand switch B7 DC output DC connection 2 1 mm only for use with the 4 Sanibel prin...

Страница 24: ...ive for upgrades USB Ports The TympStar Pro is equipped with four 4 USB ports 3 ports on the back of the instrument 2 USB Type A 1 USB Type B and one on the right side USB Type A It is possible to con...

Страница 25: ...nded to always have the USB ports enabled on the PC Disable the suspend USB power option on the PC Left Side Panel The connector to the probe box assembly is on the left side panel of the TympStar Pro...

Страница 26: ...hanical hazards only in accordance with UL 60601 1 Classified under the Medical Device Directive 93 42 EEC as a Class IIb device 0123 Conforms to European Medical Device Directive 93 42 EEC Manufactur...

Страница 27: ...GSI Part number Return to authorized representative special disposal required Manufacturer...

Страница 28: ...tor to change the ear being tested and to start and stop the test The selected ear is indicated on the screen and on the probe box as the ear that has the light There is Velcro on the back of the box...

Страница 29: ...pplication Patient The PATIENT button displays Patient screen and allows the examiner to create a new session and enter patient demographics The Patient List screen allows the user to select a patient...

Страница 30: ...tions as a toggle to switch between reflex threshold growth and reflex decay ETF The ETF key displays the Eustachian Tube Function Test screen This button functions as a toggle to switch between the I...

Страница 31: ...direction of the pressure change is indicated by the arrow The left button changes the pressure from positive to negative and the right button changes the pressure from negative to positive The press...

Страница 32: ...to allow printing of the tympanometric test results directly The printer must be HP PCL3 HP PCL5E HP PCL3 GUI or TSPL compatible If a USB memory stick is inserted into an available USB port print to P...

Страница 33: ...st options indicate the data that will be printed and the user may choose between the selected tab the marked tabs or all tabs A single ear or both ears may be selected Printout The printout settings...

Страница 34: ...btained from GSI or an authorized GSI representative A message is displayed if no USB disk is found Print Settings Selecting the button displays a printer settings dialog to select the printer protoco...

Страница 35: ...4 paper sizes 3 roll paper for the Sanibel printer and the LPQ 80 and 3 labels for the LPQ 80 Print Button Action o Screen Print prints the data displayed on the screen o Summary Print prints the mark...

Страница 36: ...ion displayed on the TympStar Pro LCD varies depending on the Test Type There are common elements found on all screens such as the Ear Tab Probe Status indicator Admittance meter Pressure meter the Te...

Страница 37: ...ters that may be changed in analysis You may change the baseline view or the immittance component on the test menu In the settings menu you may change the display options These parameter changes will...

Страница 38: ...s the settings dialog The settings dialog displays the user selectable parameters for the specific test type Comments Icon The comments icon pencil opens a comments window An external keyboard is used...

Страница 39: ...eening results are those that will be transferred to GSI Suite or printed directly from the device Only a single screening result per probe tone frequency may be marked A total of four screening resul...

Страница 40: ...ays the real time admittance as it is being recorded from the probe in the ear canal The units are millimhos mmho The current value is displayed at the top left of the meter The pressure meter display...

Страница 41: ...application TympStar Pro Config App Reflex Parameter Buttons Up to four reflexes may be recorded on the Screener display and there are two parameter buttons for each reflex For each reflex the choice...

Страница 42: ...Start check box determines if the test starts automatically upon obtaining a seal When the box is checked the tympanogram data collection begins as soon as a seal is obtained When the box is not check...

Страница 43: ...the cursor may be moved to a different point on the waveform by touching the waveform at the desired point or by using the manual pressure knob on the instrument to move right and left on the waveform...

Страница 44: ...fined in the PC Config App The device default settings can be found in the Appendix Comments The comments icon opens a comments window An external keyboard is used to enter comments Comments may be en...

Страница 45: ...m that will be transferred to GSI Suite or printed directly from the device The peak pressure of the marked tympanogram is used to set the pressure for conducting the acoustic reflex tests Only a sing...

Страница 46: ...internal error on the device Admittance and Pressure Meters The admittance meter displays the real time admittance as it is being recorded from the probe in the ear canal The units are millimhos mmho...

Страница 47: ...daPa and admittance mmho value of the indicated peak of the tympanogram The Tymp Measure t column displays the tympanometric width daPa or gradient mmho The gradient option is selected from the Setti...

Страница 48: ...button is pressed a submenu is displayed where the user may choose between the Admittance Y and the two components of Admittance Susceptance B and Conductance G and the combined graph showing both B a...

Страница 49: ...f the dialog The crosshairs of the cursor may be moved to a different point on the waveform by touching the waveform at the desired point or by using the manual pressure knob on the instrument to move...

Страница 50: ...Start Pressure The Start Pressure Test Options button is used to select the starting pressure daPa for collecting a tympanogram When the button is selected a dropdown list of starting pressures is pre...

Страница 51: ...a will display when Baseline Configuration is set to Auto or Negative Tail When Baseline Configuration is set to Positive Tail the normative data box will display only when Baseline is off NOTE Normat...

Страница 52: ...rent protocol is shown in the Protocol button When the button is pressed the list of available protocols is presented The tests included in the protocol are displayed to the right of the button The au...

Страница 53: ...reflex graph The tabs at the top of the graph are used to select the ear to be tested The area to the right of the graphic area displays the admittance and pressure meters along with the eliciting sti...

Страница 54: ...ot set NS Not set indicates no threshold was selected and the test has not been marked as no response For tests that are not set selecting the threshold or marking as a no response is required to indi...

Страница 55: ...controlled and is defined in the reflex settings dialog Reflex Test Menu Auto Zero The Auto Zero menu option automatically sets the admittance to zero when the key is pressed The system automatically...

Страница 56: ...pied from the collection area to the stored display area Only a single reflex per stimulus may be stored If you try to store a new reflex when a similar reflex has already been stored you will be prom...

Страница 57: ...lex stimulus The Mark No Response button selects the curve and marks the test as a No Response The Set Peak button sets the current point as indicated by the cross hairs as the peak of the curve The F...

Страница 58: ...n using the threshold seek function Start Level The Start Level Threshold Seek Option button is used to select the initial stimulus level dB HL for the reflex threshold testing when the threshold seek...

Страница 59: ...that indicates a reflex threshold The buttons are used to increase decrease the level When the reflex reaches the threshold testing is stopped and the level dB HL where the threshold was found and ver...

Страница 60: ...antity The Quantity Option is used to set the number of reflexes per stimulus The reflex may be repeated and all the reflexes are displayed The buttons are used to increase decrease the number of time...

Страница 61: ...l reflex decay tests The reflex collection data are displayed as graphs with different probe tones stored as tabs on the left side of the reflex graph The tabs at the top of the graph are used to sele...

Страница 62: ...the real time admittance as it is being recorded from the probe in the ear canal The units are millimhos mmho The current value is displayed at the top left of the meter The pressure meter displays t...

Страница 63: ...tomatically sets the admittance to zero when the reflex test is started and when the stimulus is presented Probe The Probe button controls the probe tone frequency used for recording the reflex The bu...

Страница 64: ...ns to scroll the waveform and change the magnification Store The Store menu option saves the selected reflex graph The graph is copied from the collection area to the stored display area Only a single...

Страница 65: ...anel button is pressed When the button is selected a dropdown menu of options is presented Level dB Step The dB Step Level Option button is used to select the level dB step when the front panel button...

Страница 66: ...cts the immittance component that is displayed on the graph The button displays a menu to select one of the following Admittance Y and the two components of Admittance Susceptance B and Conductance G...

Страница 67: ...es stored as tabs on the left side of the reflex graph The tabs at the top of the graph are used to select the ear to be tested The area to the right of the graphic area displays the admittance and pr...

Страница 68: ...mittance as it is being recorded from the probe in the ear canal The units are millimhos mmho The current value is displayed at the top left of the meter The pressure meter displays the real time pres...

Страница 69: ...he admittance to zero when the reflex latency test is started and when the stimulus is presented Probe The Probe button displays the probe tone frequency used for recording reflex thresholds Stimulus...

Страница 70: ...Zoom menu option displays the zoom dialog The zoom dialog provides a magnified view of the averaged reflex latency data and displays the time and admittance values for the selected point The dialog h...

Страница 71: ...view The close button returns to the ARLT test screen Settings The Settings menu option may be used to define the parameters used to collect the reflex latency A Settings dialog is presented with the...

Страница 72: ...from the marked points Norris et al Latency Measures of the Acoustic Reflex Audiology 13 464 469 1974 Timing The Timing buttons select the on time for the presentation of the stimulus for the reflex...

Страница 73: ...sceptance B and Conductance G Comments The comments icon opens a comments window you must use external keyboard to enter comments Comments may be entered from any test screen and it is possible to rev...

Страница 74: ...ing an intact Eustachian tube function test The collected data tympanograms are displayed with different probe tones stored as tabs on the left side of the graph The tabs at the top of the graph are u...

Страница 75: ...on the device Admittance and Pressure Meters The admittance meter displays the real time admittance as it is being recorded from the probe in the ear canal The units are millimhos mmho The current val...

Страница 76: ...Test Menu Continue The continue button starts the next tympanogram in the sequence after the initial tympanogram has been collected The instruction box indicates when it is appropriate to press the co...

Страница 77: ...ed to select the sweep rate daPa sec for the pressure change when collecting a tympanogram The button label indicates the current setting When the button is selected a submenu of options is presented...

Страница 78: ...f the dialog The crosshairs of the cursor may be moved to a different point on the waveform by touching the waveform at the desired point or by using the manual pressure knob on the instrument to move...

Страница 79: ...Pressure Test Options button is used to select the starting pressure daPa for collecting a tympanogram When the button is selected a dropdown list of options is presented Range The Pressure Range Tes...

Страница 80: ...lculated at the positive tail When Negative Tail is selected Vea for all probe tones is calculated at the negative tail Peak Indicator The Peak Indicator Display Options button determines if the tympa...

Страница 81: ...information for conducting a perforated Eustachian tube function test Multiple tests may be collected and the collected data sets are stored as tabs on the left side of the graph The tabs at the top...

Страница 82: ...Green Ready Solid Green Test in Progress Blinking Green Hold ready to continue test Solid Amber Occlusion Blinking Amber Leak or possible internal error on the device Pressure Meter The pressure mete...

Страница 83: ...limit When the menu option is selected a sub menu of choices is presented When the test starts it will increase decrease the pressure of the ear canal until the maximum level is reached Time The Time...

Страница 84: ...n Tube test Multiple tests may be collected and the collected graphs with different probe tones are stored as tabs on the left side of the graph The tabs at the top of the graph are used to select the...

Страница 85: ...ar canal The units are millimhos mmho The current value is displayed at the top left of the meter The pressure meter displays the real time pressure as it is being recorded from the probe in the ear c...

Страница 86: ...llection area to the stored display area Multiple test results per test condition may be stored A maximum of 8 stored test results may be saved Zoom The Zoom magnifying glass icon menu option displays...

Страница 87: ...option may be used to define the parameters used to collect the Patulous ETF test results A Settings dialog is presented with the parameters that are available The Patulous ETF default settings are de...

Страница 88: ...lected a dropdown list of options is presented Immittance Immittance Component The Immittance button selects the immittance component that is displayed on the graph The button toggles between the Admi...

Страница 89: ...entative if an error message occurs when accessing this test _________________________________________________________________ The Multi Hz screen presents information for conducting a single multiple...

Страница 90: ...ch the LED lights on the probe box The probe may be in the following states Indicator Lights Status Blinking Green Ready Solid Green Test in Progress Solid Green Hold ready to continue test Solid Ambe...

Страница 91: ...stic equivalent volume in cm3 The Peak column displays the pressure daPa and admittance mmho value of the indicated peak of the tympanogram Tab 2 Delta Peak Start Tab 2 contains two graphs The top gra...

Страница 92: ...daPa are also displayed Tab 3 Tympanogram at Resonant Frequency The tympanogram obtained at resonant frequency is displayed in the graph immittance component B The table on lower right side of the gra...

Страница 93: ...n 250 and 2000 Hz in 50 Hz increments Select Tymp Tab 4 and press the buttons or drag the slider bar until the desired tympanogram appears Repeat for Tab 5 and 6 if desired The Tymp Tables display the...

Страница 94: ...changed on the analysis For Multi Hz testing the baseline defaults to off Sweep Rate The Sweep Rate button is used to select the sweep rate daPa sec for the pressure change when collecting a tympanog...

Страница 95: ...eft and right arrow buttons are used to scroll the waveform The and magnifier buttons increase and decrease the magnification The crosshairs of the cursor may be moved to a different point on the wave...

Страница 96: ...o define the parameters used to collect the Multi Hz test data A Settings dialog is presented with the available parameters The Multi Hz default settings may be defined in the PC Config App The device...

Страница 97: ...range is from 400 daPa to 600 daPa Normative Data The Normative Data Display Options check box is used to display the normative data box for the 226 Hz tympanogram The tympanogram must be baseline adj...

Страница 98: ......

Страница 99: ...ts that come into direct contact with the patient are subjected to standard disinfecting procedures between patients This includes physically cleaning and using a recognized disinfectant Individual ma...

Страница 100: ...de as the ear to be tested The Velcro strip should be facing away from the body The probe box is then positioned on front of the Velcro strip Probe Insertion Ear canal examination with an illuminated...

Страница 101: ...The options for the defaults for each test type are set up in the PC configuration software TympStar Pro Config App Tympanometry 1 Select the Tympanometry mode by pressing TYMP button on the front pan...

Страница 102: ...ompletion of the sweep The user may abandon automatic testing by pressing the STOP button at any point for difficult to test patients 9 Upon completion of a pressure sweep numeric values for complianc...

Страница 103: ...lamp on the probe box and indicator on the device screen changing from blinking to steady 9 View ongoing test results on the graphic display area and on the pressure and admittance meters After presen...

Страница 104: ...ated into multiple sections The sections are presented as an expandable list on the left side of the application Subsections are displayed when the individual section is expanded The Instrument sectio...

Страница 105: ...ords may be entered under the Instrument Security sections of the Config App software Examiner Passwords are user defined and may contain any combination of lower or upper case letters and numbers The...

Страница 106: ...up here Reflex Decay The reflex decay section contains the settings for collecting acoustic reflex decay In addition to the test parameters the scaling defaults are set up here ARLT The ARLT section c...

Страница 107: ...A set of tests that may be run in a sequence are setup in the Auto Sequence protocol The tests in the protocol may include diagnostic tests...

Страница 108: ...ument Services USB that was included in the original shipment of the TympStar Pro or contact your GSI representative Data Port Direct The Data Port provides backwards compatibility with the TympStar s...

Страница 109: ...ment fitting Noah 4 may be installed as standalone software or on a network Data transfer and storage utilizes the Noah database for data management AudBase AudBase software saves audiometric data fro...

Страница 110: ...repairs be performed by a qualified GSI service representative only Malfunctions resulting from improper maintenance or repair by anyone other than an authorized GSI representative will be the respons...

Страница 111: ...panogram should be identical other than the Vea should read 2 0 cm3 The same sequence may be followed with the 5 0cm3 opening on the test cavity Keep a copy of the tests for a day to day reference in...

Страница 112: ...mmended that the instrument is cleaned at least annually Grason Stadler upon request will provide circuit diagrams component part lists descriptions calibration instructions or other information to as...

Страница 113: ...up over time causing dull screens Never spray or otherwise apply any solution directly to your display It could leak into the device and damage electronic parts Cleaning and Disinfecting Agents Accor...

Страница 114: ...ased cleaners Gently wipe the items with the slightly damp cloth taking care not to get moisture in the speaker portion of the probe and contralateral earphones The eartips for the probe are for singl...

Страница 115: ...2 Take off the probe tip Note the gasket inside the tip 3 Thread the stiff end of the cleaning brush into one of the tubes from inside...

Страница 116: ...each of the four tubes Discard floss after use NOTE Only insert the thread from the inside out this will insure that the dirt is pushed out of the probe instead of into the probe and as well protect...

Страница 117: ...ool is useful for removal of cerumen that was not removed using the floss The wire on this tool can assist in pushing out the cerumen from each of the four tubes The metal loop on the end of this 3 in...

Страница 118: ...e outside for each the four tubes of the probe tip Do not use the brush end of this tool to clean the probe tip instead use the cleaning floss 3 After the cerumen has been removed reassemble the probe...

Страница 119: ...1500 Altitude Units Feet Default Ear Right Initial Screen Tympanometry Contralateral Ear Side Probe Contralateral Transducer Insert Tympanometry Types Y A A d A dd A s B C D E Tympanometry Types B G...

Страница 120: ...r Off Reflex Settings Probe Tone 226 Hz Immittance Component Y Reflex Ear Ipsi Reflex Frequency 500 Hz Reflex Stimulus Type Pulsed Reflex Averages 1 Reflex Stimulus Quantity 1 Time Base 15 sec Stimulu...

Страница 121: ...Start Off Baseline Off Sweep Rate 50 daPa sec Age Adult Normative Data 226 Hz Tab 1 Off Peak Indicator On Pressure Range Normal Start Pressure 200 daPa Immittance B Delta Peak Start Resonant frequency...

Страница 122: ...ay 1 Tymp Probe 226 2 Reflex Ipsi 500 Pulsed Off 3 Reflex Ipsi 1000 Pulsed Off 4 Reflex Ipsi 2000 Pulsed Off 5 Reflex Ipsi 4000 Pulsed Off 6 Reflex Contra 500 Pulsed On 7 Reflex Contra 1000 Pulsed On...

Страница 123: ...of Measurement Tymp mode at 226 Hz Probe Tone 0 1 mmho or 5 whichever is greater Reflex mode at 226 Hz Probe Tone 0 02 mmho or 5 whichever is greater Signal level accuracy 1 5 dB SPL Temporal Characte...

Страница 124: ...Probe RETSPL Contra Insert RETSPL Contra Headset 250 24 5 14 0 27 0 500 9 5 5 5 13 0 1000 6 5 0 0 6 0 2000 12 3 0 8 0 4000 3 5 5 5 9 0 Broad Band Noise 7 0 7 0 7 0 Low Band Noise 9 0 9 0 9 0 High Band...

Страница 125: ...l lose their calibration The frequency of the external stimulus must be carefully selected for the probe tone with which it will be used If the stimulus is too close to the probe tone frequency a stim...

Страница 126: ...ins Operation Input Voltage 100 240 VAC Input Frequency 50 60 Hz Power Consumption 60 Watts Maximum Average Operating Current Idle 220 mA Average Operating Current Running Test 290 mA Heat Generation...

Страница 127: ...Depth 16 5 cm 6 5 in Weight 5 7 kg 12 6 lb Physical Instrument Height 37 cm 14 5 in Width 41 cm 16 in Depth 28 cm 11 in Weight 5 5 kg 12 lb Probe Shoulder Box Height 9 9 cm 3 875 in Width 6 4 cm 2 5...

Страница 128: ...d the curve in the scratch pad is selected then stored as a threshold To select a curve on the scratchpad the curve is touched on the LCD or clicked with the mouse If there is no curve selected and th...

Страница 129: ...ctromagnetic Compatibility Although the instrument fulfils the relevant EMC requirements precautions should be taken to avoid unnecessary exposure to electromagnetic fields e g from mobile phones etc...

Страница 130: ...which radiated RF disturbances are controlled The customer or the user of the TympStar Pro can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile...

Страница 131: ...ferential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial hospital or residential environment Voltage Dips Short Interruptions...

Страница 132: ...d RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked Note 1...

Страница 133: ...equirements as specified in IEC 60601 1 2 it is essential to use only the following accessories Item Manufacturer Item TympStar Pro Probe Grason Stadler 8104117 IP30 Contra earphone RadioEar 8101894 D...

Страница 134: ...1987 Specification for Instruments to measure Aural Acoustic Impedance and Admittance Aural Acoustic Immittance 2 IEC 60645 5 2004 Electroacoustics Audiometric Equipment Instruments for the measuremen...

Отзывы: