ETF Perforated Test Results
The lower right side of the ETF Perforated screen displays the test results in a
table. The header of the table indicates the test number. The Result column lists
the opening/closing row number. Each time the instrument detects and opening
and closing of the Eustachian tube, the results are recorded in the appropriate
columns. The time and pressure of the detected opening/closing is stored in the
table
ETF Perforated Test Menu
Max Pressure
The Max Pressure button is used to select the maximum pressure limit. When
the menu option is selected, a sub menu of choices is presented. When the test
starts it will increase/decrease the pressure of the ear canal until the maximum
level is reached.
Time
The Time button sets the length of the test in seconds. When the menu option is
selected, a sub menu of choices is presented.
Mark Test
The Mark Test menu option is used to indicate the ETF perforated tab that will
be used when the data is transferred to GSI Suite. The marked tab is indicated by
the check mark in the tab area.
Comments
The comments icon opens a comments window (you must use external keyboard
to enter comments). Comments may be entered from any test screen and it is
possible to review and edit comments from any test screen.